Home » Electronic Measuring Instruments » Optical, Light & Color Measurement » Colorimeter » TechnoOptis UA-10A Luminance & Chromaticity Uniformity Analyzer

TechnoOptis UA-10A Luminance & Chromaticity Uniformity Analyzer

TechnoOptis UA-10A Luminance and Chromaticity Uniformity Analyzer

Fast two-dimensional luminance and color uniformity analysis for automotive displays, LED backlights and automated optical production inspection.

The TechnoOptis UA-10A is a compact CMOS imaging colorimeter used to map luminance and chromaticity across a display or illuminated surface. Its 1920 × 1200 measurement-point array supports visual inspection of spatial nonuniformity, display Mura and light-distribution patterns that are difficult to characterize with a single-point meter.

Standard UA-10A configurations measure luminance from 0.1 to 80,000 cd/m². Optional neutral-density filters expand the available high-luminance applications; the manufacturer quotes 10 to 800 million cd/m² for the supported ND-filter configuration, not as a standard filter-free range.

Choose the UA-10AS standard lens, UA-10AW wide-angle lens or UA-10AT telephoto lens to match field of view to the target and installation distance. The Gigabit Ethernet interface and supplied external-control SDK support integration into production inspection stations and automated equipment.

Manufacturer-published best-case measurement time is approximately 0.98 second and varies with use conditions. The CI-100 measurement and analysis software provides display modes, result visualization and optical uniformity analysis.

2.3 MP CMOS 1920 × 1200 spatial measurement points
0.1–80,000 cd/m² Standard luminance measurement range
Approx. 0.98 s Fastest stated measurement time; condition-dependent
3 optical versions Standard, wide-angle and telephoto

2.3-Megapixel Spatial Measurement

A 1920 × 1200 CMOS image sensor records point-by-point luminance and chromaticity for detailed display uniformity maps.

Fast Inspection Cycle

Manufacturer-stated fastest measurement is about 0.98 second; actual timing depends on optics, exposure and operating conditions.

Three Field-of-View Options

Choose standard, wide-angle or telephoto versions rather than assuming one instrument covers all viewing geometries.

Wide Luminance Capability

Standard 0.1–80,000 cd/m² operation extends to brighter devices with the correct optional neutral-density filter.

Automated Station Integration

Gigabit Ethernet communication and an external-control SDK facilitate programmable inspection sequences and data collection.

CI-100 Uniformity Analysis

Compatible measurement and analysis software provides spatial displays, evaluation functions and reporting support.

Automotive Instrument Clusters

Map brightness and color nonuniformity in instrument panels, interior illumination and illuminated controls.

LCD and OLED Display QC

Inspect spatial luminance and chromaticity Mura in panels, assemblies and related optical components.

Micro LED and Mini LED Development

Characterize brightness and color distributions of high-brightness emissive devices and backlights.

AR/VR Optical Inspection

Evaluate luminance and color uniformity over areas relevant to near-eye display modules.

LED Lighting and Light Guides

Examine luminous surfaces, illuminated panels and backlight distribution for uneven output.

Production Line Integration

Deploy automated 2D optical inspection for incoming checks, process tuning and end-of-line quality screening.

UA-10A Specifications

Model familyUA-10A
Detector2.3-megapixel CMOS image sensor
Measurement points1920 × 1200
Standard luminance range0.1–80,000 cd/m²
Optional ND-filter luminance range10–800,000,000 cd/m²; proper filter required
Luminance linearity±2% under manufacturer reference conditions
Chromaticity accuracy±0.003 under manufacturer reference conditions
In-plane luminance uniformity±2% under manufacturer reference conditions
In-plane chromaticity uniformity±0.003 under manufacturer reference conditions
Luminance repeat characteristic0.4% (2σ)
Chromaticity repeat characteristic0.002 (max–min)
Fastest stated measurement timeApproximately 0.98 second; use-condition dependent
Luminance stability±1%
Luminance repeatability±2%
Temperature influenceLuminance ±3% relative to 20°C over stated 0–40°C condition
Humidity influenceLuminance ±3%, up to 80% RH noncondensing
CommunicationGigabit Ethernet (GigE)
Control integrationExternal-control SDK
Analysis softwareCI-100
Supported PC OSWindows 10 Pro / Windows 11 Pro, 64-bit, per manufacturer documentation
Recommended PC processorIntel Core i7-11700 or higher
Recommended PC memory16 GB or higher
Recommended storage500 GB or higher; at least 3 GB free system-disk space
Product classification2D luminance and chromaticity imaging colorimeter

Model and Lens Configuration

UA-10AS | Standardf = 12 mm; 131.2 × 69.3 × 61.0 mm; approximately 510 g
UA-10AW | Wide-anglef = 6.5 mm; 140.9 × 69.3 × 64.0 mm; approximately 580 g
UA-10AT | Telephotof = 50 mm; 131.3 × 69.3 × 59.5 mm; approximately 490 g
At 1,000 mm distanceUA-10AS: 496.8 × 310.5 mm; UA-10AW: 917.2 × 573.2 mm; UA-10AT: 121.7 × 76.0 mm (manufacturer table, 90% FOV)

Reference accuracy conditions: standard illuminant A; color and luminance accuracy evaluated at the center of the CMOS. Do not treat the optional ND-filter rating as the normal range.

Manufacturer tables use both UA-10A and UA-10AS for the standard-lens configuration; confirm final order code.