Home » Electronic Measuring Instruments » Optical, Light & Color Measurement » Colorimeter » TechnoOptis RD-80SA Luminance Colorimeter

TechnoOptis RD-80SA Luminance Colorimeter

RD-80SA High-Speed Luminance Colorimeter

The TechnoOptis RD-80SA combines spot luminance and chromaticity measurement with fast analog output for evaluating LCD response characteristics and light-source flicker.

Its photomultiplier-tube detector measures luminance from 0.1 to 10,000 cd/m², while the analog output responds in 80 µs or less under the manufacturer’s specified test condition. This 80 µs figure is an analog response characteristic—not the time for one complete digital measurement, which is approximately one second under stated conditions.

Engineers can examine Gray-to-Gray transitions, overshoot, undershoot, and PWM-modulated light output using the analog waveform with an external oscilloscope. The RD-80SA also reports spot luminance, chromaticity, and related optical parameters for LCD panels, LED backlights, and lighting evaluation.

≤ 80 µs Analog output response
0.1–10,000 cd/m² Luminance measurement range
0.04 cd/m² Small-flicker detection capability
RS-232C / LAN Digital interfaces

Fast analog response

Analog output reaches its specified response in 80 µs or less, supporting waveform-based evaluation of rapidly changing luminance.

Photomultiplier sensitivity

Photomultiplier-tube detection supports subtle changes in optical output, including reported flicker components as small as 0.04 cd/m².

Spot luminance and color

A 2° measurement angle provides targeted luminance and chromaticity data for a selected area.

Eight luminance ranges

Automatic or manual range selection covers measurements from 0.1 to 10,000 cd/m².

Waveform observation

The approximately 0 to +3.4 V analog output can be viewed with external recording or oscilloscope equipment.

System connectivity

RS-232C and LAN interfaces support instrument integration into optical test workflows.

LCD Gray-to-Gray Testing

Capture transitions in screen luminance to evaluate response behavior and motion-related artifacts.

PWM Backlight Analysis

Inspect brightness modulation produced by pulse-width-modulated LED backlights.

Flicker Characterization

Detect low-amplitude temporal variations in display panels and illumination systems.

Overdrive Circuit Assessment

Investigate luminance overshoot and undershoot during dynamic display transitions.

Optical R&D

Compare luminance and chromaticity behavior across display materials and driving conditions.

Automated Display Inspection

Integrate spot optical measurement and communications into repeatable inspection setups.

Key specifications

Parameter Value
Product / detector RD-80SA / photomultiplier tube
Measurement type Spot luminance, chromaticity, and analog temporal light response
Optical system Objective lens f=80 mm, F2.5; 5° eyepiece field
Measurement angle 2°
Measurement distance 350 mm to infinity (from objective-lens metal tip)
Luminance range 0.1–10,000 cd/m²; eight auto/manual ranges
Luminance accuracy ±3% at ≤2 cd/m²; ±2% above 2 cd/m² (standard illuminant A, auto range)
Chromaticity accuracy (x,y) ±0.0040 at ≤2 cd/m²; ±0.0025 above 2 cd/m² (standard illuminant A, auto range)
Analog output response ≤80 µs; 90% response to a pulsed LED under the stated test condition
Approximate digital measurement time 1.0 s (SINGLE mode, Y filter, Range 4, automatic calibration off)
Analog output Approximately 0 to +3.4 V
Interfaces RS-232C; LAN
Power supply / consumption 100–240 V AC adapter, 50/60 Hz; approximately 34 W
Operating environment 0–35°C; ≤65% RH, non-condensing
Dimensions / weight 130 × 319 × 201 mm (W×D×H); 4.3 kg

Luminance range by selected instrument range

Parameter Range 1 Range 2 Range 3 Range 4 Range 5 Range 6 Range 7 Range 8
Nominal luminance span 0.1–5 0.5–15 1.5–40 3.5–120 15–600 35–1,600 220–2,900 750–10,000

All photometric accuracy values are for the manufacturer’s reference conditions. The 80 µs specification is the analog-output response, not the normal digital measurement cycle. External waveform instrumentation is required for time-domain analysis; nominal range boundaries are design references.