TechnoOptis SR-5S Spectroradiometer

TechnoOptis SR-5S Spectroradiometer

Visible spectral radiance and precise color measurement for display modules, instrument panels and lighting quality control.

The TechnoOptis SR-5S is a spectroradiometer intended for optical-color evaluation of flat-panel displays, automotive instrument panels, traffic signals and specialty lighting. The diffraction-grating and cooled-CCD design provides wavelength-resolved measurement from 380 to 780 nm.

Four motorized measuring angles (2°, 1°, 0.2°, 0.1°) offer flexibility in spot size, while the 2° aperture supports luminance as low as 0.001 cd/m². The maximum published luminance of 500,000 cd/m² is at 0.1°; other angles have different limits.

TechnoOptis offers an enhanced chromaticity-accuracy configuration with x ±0.0015 and y ±0.001. The manufacturer explicitly identifies this as a custom specification, so it should not be promised as standard for all SR-5S orders.

A touch panel, USB 3.0 and RS-232C interface support practical instrument operation and data collection. For HDR ranges beyond the standard SR-5S class, consider the SR-5A or the higher-sensitivity SR-5AS.

0.001 cd/m² 2° measurement lower limit
380–780 nm Visible-range spectral analysis
4 angles 2° / 1° / 0.2° / 0.1°
Custom xy x ±0.0015 / y ±0.001 option

Spectral Color Measurement

Capture spectral radiance and derive CIE chromaticity, tristimulus XYZ and correlated color temperature.

Low-Luminance Measurement

Analyze dim targets down to 0.001 cd/m² under the specified 2° conditions.

Custom High-Chromaticity Accuracy

Order an enhanced xy accuracy specification when precise color acceptance limits require it.

Motorized Measuring Angles

Select four fields of view to suit small display regions and different target sizes.

High-Speed Measurement Mode

Reduce measurement time with the instrument’s supported rapid acquisition mode.

Touchscreen and Digital Interfaces

Operate via touchscreen and connect through RS-232C or USB 3.0 for supported workflows.

Flat-Panel Display Color Testing

Compare spectral and chromaticity output of LCD, OLED and related materials.

Automotive Instrument Lighting

Measure dashboard indicators, control panels and interior display luminance and color.

LED and Specialty Lighting

Check visible spectrum and correlated color temperature of light sources.

Traffic Signals and Signage

Evaluate signal lamp color coordinates and visibility-related brightness characteristics.

Production Color Acceptance

Record calibrated colorimetric data for display assembly quality checks.

Optical Materials R&D

Characterize light-emitting materials and compare emission characteristics among prototypes.

SR-5S Specifications

ModelSR-5S
DetectorElectronically cooled linear CCD
Dispersive elementDiffraction grating
Objective / viewfinderf = 82 mm, F2.5 objective; 5° viewfinder field
Measuring angles2° / 1° / 0.2° / 0.1°; motorized selection
Spectral measurement range380–780 nm
Wavelength resolution1 nm
Spectral wavelength accuracy±0.3 nm, measured with mercury emission lamp
Spectral measurement quantitySpectral radiance (W·sr⁻¹·m⁻²·nm⁻¹)
Photometric/colorimetric measurementsRadiance, luminance, tristimulus XYZ, CIE 1931 xy, CIE 1976 u′v′, correlated color temperature/duv, and spectral peak; 2°/10° standard observer
Reference luminance accuracy±2% or less, under manufacturer reference conditions
Polarization sensitivityLuminance ≤1%; spectral radiance ≤2% over 400–780 nm
SupplyDedicated AC adapter; 100–240 V AC, 50/60 Hz; 12 V DC instrument input
WeightApprox. 5.5 kg
Measuring distance250 mm to infinity, from objective-lens metal tip
Spectral bandwidth5 nm FWHM
Luminance range — 2°0.001–15,000 cd/m²
Luminance range — 1°0.003–45,000 cd/m²
Luminance range — 0.2°0.075–125,000 cd/m²
Luminance range — 0.1°0.3–500,000 cd/m²
Enhanced chromaticity accuracyx ±0.0015, y ±0.001 — custom configuration, not universal standard
Luminance repeatability0.7% from 0.001–0.1 cd/m²; 0.3% at ≥0.1 cd/m² (specified 2°/normal-mode conditions)
Chromaticity repeatability0.005 from 0.001–0.1 cd/m²; 0.0005 at ≥0.1 cd/m² (manufacturer test conditions)
Measurement modesAuto/manual integration; high-speed mode; vertical-sync input shown in published specifications
Digital interfacesRS-232C and USB 3.0
Power consumptionApprox. 30 W
Operating conditions0 to 35°C; ≤80% RH, noncondensing
Dimensions (W × D × H)422 × 130 × 244 mm

Measurement area by field angle

Angle250 mm350 mm500 mm1,000 mm
2°6.5 mm10.0 mm15.1 mm32.2 mm
1°3.25 mm4.99 mm7.55 mm16.1 mm
0.2°0.65 mm1.00 mm1.51 mm3.22 mm
0.1°0.33 mm0.50 mm0.76 mm1.61 mm

Measurement distances are from the objective-lens metal fixture tip; spot diameters shown in the table are manufacturer reference data. Ranges/accuracy are conditional on aperture, light source and test conditions.

Enhanced chromaticity accuracy is labeled a custom-made model on the TechnoOptis product page. Obtain quotation for the chosen accuracy option rather than representing it as a standard specification.

Measurable luminance is angle dependent; the overall 0.001–500,000 cd/m² endpoints occur at different measuring angles.

All measuring-range and accuracy specifications are subject to manufacturer reference conditions.