TechnoOptis SR-UL1R Spectroradiometer

SR-UL1R Ultra-Low-Luminance Spectroradiometer

The discontinued TechnoOptis SR-UL1R is a visible-range spectroradiometer developed for luminance and color evaluation of dim displays, instrumentation panels, and optical devices.

At its 2° observation angle, the SR-UL1R measures luminance from 0.001 to 3,000 cd/m². Its overall family of angle-dependent ranges extends to 300,000 cd/m², but minimum and maximum levels change with the selected motorized angle.

The SR-UL1R covers 380–780 nm with a 1 nm wavelength resolution, cooled linear CCD, and motor-selectable 2°, 1°, 0.2°, or 0.1° spot measurement. USB/RS-232C support and the manufacturer’s CS-900A software facilitate measurement control and review.

Production has ended. The manufacturer lists SR-5 as its successor; retain this page for legacy instrument comparison and replacement evaluation.

0.001 cd/m² Minimum at 2°
380–780 nm Spectral range
4 angles 2° / 1° / 0.2° / 0.1°
Discontinued Manufacturer successor: SR-5

Low-luminance measurement

The 2° setting begins at 0.001 cd/m², supporting near-black display testing.

Angle-specific dynamic range

Covers up to 300,000 cd/m² at the 0.1° angle; minimum thresholds vary by aperture.

Visible spectral analysis

Provides 380–780 nm wavelength coverage with 1 nm resolution.

Motorized angle switching

Four selectable apertures accommodate display areas and different spot sizes.

Synchronization support

External vertical synchronization input supports tests of periodically driven light sources.

Legacy PC control

USB 2.0, RS-232C, and CS-900A support measurements in existing test environments.

OLED Dark-Level Verification

Assess very low luminance and chromaticity during display development.

LCD Optical Characterization

Measure point luminance and chromaticity across display test samples.

Automotive Instrument Lighting

Evaluate dim instrument clusters, control panels, and cabin displays.

Digital Signage Inspection

Inspect visible spectral properties of display sections.

Traffic Signals and Indicators

Characterize luminous color and intensity of signaling products.

Legacy Instrument Qualification

Document capabilities and replacement requirements for existing SR-UL1R units.

Product statusDiscontinued; manufacturer successor SR-5
Instrument typeSpot spectroradiometer
Wavelength range380–780 nm
Spectral half-bandwidth5–8 nm
Wavelength resolution1 nm
Wavelength accuracy±0.3 nm, Hg lamp reference
Photo detectorElectronically cooled linear CCD
Measurement angles2°, 1°, 0.2°, 0.1°; motor selectable
Measurement distance350 mm to infinity
Luminance accuracy±2% or less, manufacturer reference conditions
Chromaticity accuracy±0.002, manufacturer reference conditions
InterfacesUSB 2.0 / RS-232C
Measurement modesNORMAL, HIGH SPEED; auto/manual
Power consumptionApprox. 36 W
Dimensions (W × D × H)150 × 406 × 239 mm
WeightApprox. 5.5 kg
Measurement angle2°1°0.2°0.1°
Luminance range (cd/m²)0.001–3,0000.003–9,0000.075–70,0000.3–300,000

The 0.001 cd/m² minimum applies at 2° only. The 300,000 cd/m² maximum applies at 0.1°.

Manufacturer source: https://www.techno-optis.com/en/products/spectroradiometer-sr-ul1r/