TechnoOptis SR-3AR Spectroradiometer

SR-3AR Spectroradiometer (Discontinued)

The TechnoOptis SR-3AR is a discontinued spot spectroradiometer designed for high-accuracy luminance, chromaticity, and spectral radiance analysis of displays and other light-emitting products.

This legacy model uses an electronically cooled linear CCD and diffraction grating to evaluate wavelengths from 380 to 780 nm at 1 nm sampling resolution. Motorized selection of 2°, 1°, 0.2°, or 0.1° measurement angles helps isolate small targets without an attachment lens under suitable conditions.

The SR-3AR is retained as a reference for users maintaining existing test programs. TechnoOptis identifies the SR-5 as its successor; for new purchasing or replacements, request a current-model comparison rather than assuming the SR-3AR remains available.

Discontinued Legacy product status
380–780 nm Spectral coverage
0.1°–2° Selectable measurement angles
SR-5 Manufacturer-listed successor

Spectral measurement

Diffraction-grating optics and an electronically cooled linear CCD support spectral radiance analysis.

Four measurement angles

Motorized selection of 2°, 1°, 0.2°, or 0.1° allows spot measurement of differently sized targets.

Visible-range coverage

Covers 380–780 nm with 1 nm wavelength resolution for display and lighting color analysis.

Photometric and color data

Reports luminance, chromaticity coordinates, tristimulus values, and correlated color temperature.

PC interfaces

USB 2.0 and selectable RS-232C enable connection to existing measurement systems.

Legacy software support

The original package included CS-900A application software for control and evaluation.

Legacy Display Test Systems

Maintain historical test methods and data continuity for LCD and other flat-panel display platforms.

Automotive Instrument Panels

Evaluate spectra and chromaticity of gauges, indicators, and illuminated controls.

Small-Area Color Measurement

Inspect compact luminous targets using the instrument’s narrower measurement angles.

Traffic Signals

Characterize luminance and colorimetric performance of traffic lights and information signs.

Digital Signage

Study emissive color and luminance in outdoor and indoor signage.

Instrument Replacement Planning

Review legacy SR-3AR requirements when specifying a current SR-5 replacement.

Key specifications

Parameter Value
Product status Discontinued; manufacturer-listed successor SR-5
Detector Electronically cooled linear CCD
Spectral system Diffraction grating
Wavelength range / resolution 380–780 nm / 1 nm
Spectral bandwidth 5–8 nm (half bandwidth)
Wavelength accuracy ±0.3 nm (Hg emission lamp)
Measurement angles 2°, 1°, 0.2°, 0.1°; motorized selection
Measurement distance 350 mm to infinity
Luminance range at 2° 0.1–3,000 cd/m²
Luminance range at 1° 0.3–9,000 cd/m²
Luminance range at 0.2° 7.5–70,000 cd/m²
Luminance range at 0.1° 30–300,000 cd/m²
Luminance accuracy ≤±2% under manufacturer reference conditions
Chromaticity accuracy (x,y) ±0.002 under reference conditions
Interface USB 2.0 or RS-232C
Power / consumption 100–240 V AC adapter; approximately 34 W
Dimensions / weight 150 × 406 × 239 mm (W×D×H); approximately 5.5 kg

Original spot-size and brightness ranges

Parameter 2° 1° 0.2° 0.1°
Minimum measurement area at 350 mm 10.0 mm Ø 4.99 mm Ø 1.0 mm Ø 0.5 mm Ø
Luminance span (cd/m²) 0.1–3,000 0.3–9,000 7.5–70,000 30–300,000

Historical SR-3AR specifications are provided for installed equipment and model comparison, not as a statement of current product availability. Luminance range is measurement-angle-dependent; accuracy assumes manufacturer test conditions. SR-5 specifications differ and must not be substituted for SR-3AR values.