TechnoOptis SR-LEDW Spectroradiometer

TechnoOptis SR-LEDW Spectroradiometer

Discontinued legacy spectroradiometer for LED, display and low-light spectral measurement; upgrade path: TechnoOptis SR-5A.

The TechnoOptis SR-LEDW is a discontinued spectroradiometer previously used for very-low-luminance display measurement, high-brightness LED evaluation and spectral-color analysis. Its original visible-range sensor captured spectral radiance over 380–780 nm with a cooled linear CCD and diffraction grating.

The published aggregate luminance span is 0.0005–5,000,000 cd/m², but this combines different measurement angles. The 0.0005 cd/m² lower endpoint belongs to the 2° aperture, while 5,000,000 cd/m² belongs to 0.2°. At 0.1°, the published upper limit was 2,000,000 cd/m².

The legacy FIX mode was designed to shorten repeated measurements on the same target by optimizing measurement settings. An illuminance adapter was available separately. These features are retained here as historical reference information, not as claims of current new-unit availability.

TechnoOptis identifies the SR-5A as the successor model. Engineers replacing an SR-LEDW should compare aperture-specific luminance ranges, communication interfaces, working distances and software integration before selecting a replacement.

Discontinued Manufacturer-confirmed product status
380–780 nm Legacy visible-range measurement
5 M cd/m² Maximum at 0.2° measuring angle
SR-5A Manufacturer-listed replacement

Legacy Wide Dynamic Range

Measured low display light and high-brightness LEDs across angle-specific luminance ranges.

Cooled-CCD Optical Spectroscopy

Acquired visible-range spectral radiance via a diffraction-grating design.

Four Motorized Apertures

Selectable 2°, 1°, 0.2° and 0.1° fields of view for different target sizes.

FIX Repeated-Measurement Mode

Optimized integration settings for repeated readings of the same target.

CIE Color and Luminance Analysis

Provided luminance, radiance, chromaticity coordinates, tristimulus XYZ and correlated color temperature.

Confirmed SR-5A Upgrade Path

Use the manufacturer-recommended replacement for new projects and compare interface/fixture compatibility.

Existing SR-LEDW Installations

Reference original specifications when maintaining established optical test stations.

LED Chip and Module Development

Review legacy spectral and luminance readings on LED emitters.

Display Black-Level Characterization

Interpret earlier low-luminance measurements on backlights and panels.

LCD and OLED Color Evaluation

Reference spectral-color test methods used with the legacy instrument.

Automotive Lighting Comparison

Maintain historical test baselines for illuminated displays and headlamps.

Instrument Replacement Planning

Translate SR-LEDW measurement procedures to the current SR-5A with application-specific validation.

SR-LEDW Specifications

ModelSR-LEDW
DetectorElectronically cooled linear CCD
Dispersive elementDiffraction grating
Objective / viewfinderf = 82 mm, F2.5 objective; 5° viewfinder field
Measuring angles2° / 1° / 0.2° / 0.1°; motorized selection
Spectral measurement range380–780 nm
Wavelength resolution1 nm
Spectral wavelength accuracy±0.3 nm, measured with mercury emission lamp
Spectral measurement quantitySpectral radiance (W·sr⁻¹·m⁻²·nm⁻¹)
Photometric/colorimetric measurementsRadiance, luminance, tristimulus XYZ, CIE 1931 xy, CIE 1976 u′v′, correlated color temperature/duv, and spectral peak; 2°/10° standard observer
Reference luminance accuracy±2% or less, under manufacturer reference conditions
Polarization sensitivityLuminance ≤1%; spectral radiance ≤2% over 400–780 nm
SupplyDedicated AC adapter, 100–240 V AC, 50/60 Hz; instrument DC 12 V
WeightApprox. 5.5 kg
Product lifecycleDiscontinued; SR-5A listed as successor by TechnoOptis
Measuring distance350 mm to infinity, from objective-lens metal tip
Spectral bandwidth5–8 nm half-bandwidth
Luminance range — 2°0.0005–1,500,000 cd/m²
Luminance range — 1°0.0015–4,500,000 cd/m²
Luminance range — 0.2°0.0375–5,000,000 cd/m²
Luminance range — 0.1°0.15–2,000,000 cd/m²
Chromaticity accuracy±0.002 under specified manufacturer reference conditions
Luminance repeatability1.5%: 0.0005–0.005 cd/m²; 0.4%: 0.005–0.1 cd/m²; 0.3%: ≥0.1 cd/m²
Chromaticity repeatability0.005: 0.0005–0.005 cd/m²; 0.0015: 0.005–0.1 cd/m²; 0.0005: ≥0.1 cd/m²
Measurement modesAuto/manual, legacy FIX mode, vertical synchronization input
Measurement time — normalApprox. 1–248 s, dependent on measurement conditions
Measurement time — high-speedApprox. 1–17 s, dependent on measurement conditions
Digital interfacesUSB 2.0 / RS-232C selectable
Power consumptionApprox. 36 W
Operating conditions5 to 30°C; ≤80% RH, noncondensing
Dimensions (W × D × H)150 × 406 × 239 mm

Measurement area by field angle

Angle350 mm500 mm800 mm1,000 mm
2°10.0 mm15.1 mm25.4 mm32.2 mm
1°4.99 mm7.55 mm12.7 mm16.1 mm
0.2°1.00 mm1.51 mm2.54 mm3.22 mm
0.1°0.50 mm0.76 mm1.27 mm1.61 mm

Measurement distances are from the objective-lens metal fixture tip; spot diameters shown in the table are manufacturer reference data. Ranges/accuracy are conditional on aperture, light source and test conditions.

The SR-LEDW is officially discontinued. Do not imply new inventory, price or immediate delivery without distributor confirmation.

The 0.0005–5,000,000 cd/m² aggregate span cannot be achieved at one aperture; angle-specific ranges must be used in customer specifications.

The successor SR-5A has a shorter minimum working distance (250 mm vs 350 mm), different physical dimensions, and USB 3.0 versus legacy USB 2.0; integration is not automatically plug-and-play.