Single-Phase Harmonic and Flicker Compliance Testing

The KIKUSUI KHA1000 is a harmonic and flicker analyzer dedicated to single-phase, two-wire electrical equipment, combining standards-based harmonic-current and voltage-fluctuation testing with general power and waveform measurement functions.

The analyzer provides one voltage and one current measurement channel, voltage ranges of 150 V and 300 V, and six current ranges from 0.5 A to 20 A. It is designed primarily for harmonic-current and flicker compliance testing of single-phase equipment within the applicable 16 A standards range.

Single-Phase 2-Wire Dedicated architecture for single-phase equipment testing
IEC 61000-3-2 Harmonic-current compliance testing for applicable equipment
IEC 61000-3-3 Flicker and voltage-fluctuation evaluation
No PC Required Test setup, judgment, and report creation on the instrument

Beyond formal compliance testing, the KHA1000 measures voltage, current, active power, apparent power, reactive power, power factor, frequency, and phase while also providing waveform monitoring, FFT analysis, and inrush-current capture. GPIB, RS232C, and USB interfaces are standard for automated measurement environments.

Dedicated Single-Phase Compliance Testing

The KHA1000 is designed specifically for single-phase, two-wire equipment, providing a focused platform for IEC 61000-3-2 harmonic-current and IEC 61000-3-3 flicker and voltage-fluctuation testing.

Harmonic Analysis up to the 180th Order

Harmonic analysis extends through the 40th order in standards-oriented HA mode and through the 180th order in OTHER mode for broader development and diagnostic measurements.

Interharmonic Measurement Support

Selectable processing supports IEC 61000-4-7 measurement methods with interharmonic processing enabled or disabled according to the selected measurement-standard configuration.

Flicker and Voltage-Fluctuation Analysis

Built-in flicker analysis measures Pst and voltage-fluctuation parameters, with selectable simultaneous or independent voltage-fluctuation measurement and support for manually switched equipment.

FFT and Inrush-Current Measurement

Waveform monitoring and FFT analysis support development work, while direct inrush-current measurement captures peaks up to 80 Apeak for up to 20 ms without requiring a separate oscilloscope for basic inrush evaluation.

Standalone Test and Report Workflow

Test conditions, execution, standards-based pass/fail judgment, spectrum review, and result-report generation can be performed directly from the KHA1000 without a PC for normal compliance-test workflows.

Single-phase electrical appliance harmonic-current testing
Consumer and office electronics harmonic compliance evaluation
Single-phase power supply and electronic equipment emission testing
IEC 61000-3-2 harmonic-current compliance testing
IEC 61000-3-3 flicker and voltage-fluctuation testing
Pre-compliance waveform, FFT, and inrush-current evaluation

KHA1000 Specifications

Measurement Configuration1 voltage channel / 1 current channel
Supported WiringSingle-phase 2-wire
Maximum Input Voltage300 Vrms / 560 Vpeak (CAT I), 250 Vrms (CAT II)
Voltage Ranges150 V / 300 V
Voltage MeasurementTrue RMS and ±peak
Voltage Accuracy±(0.4% of reading + 0.04% of range)
Maximum Input Current24 Arms / 50 Apeak
Short-Duration Peak Current80 Apeak within 20 ms
Current Ranges0.5 A / 1 A / 2 A / 5 A / 10 A / 20 A
Current Accuracy — 0.5 A Range±(0.5% of reading + 0.2% of range), 45 Hz to 65 Hz
Current Accuracy — 1 A to 20 A±(0.5% of reading + 0.1% of range), 45 Hz to 65 Hz
Frequency Measurement Range45 Hz to 65 Hz
Frequency Resolution0.001 Hz
Power MeasurementsActive power, apparent power, reactive power, and power factor
Phase Range / Resolution0.00° to 359.99° / 0.01°
Harmonic Analysis40th order in HA mode / 180th order in OTHER mode
Voltage Harmonic AnalysisThrough the 40th order
Pst Accuracy1 ±5%
Pst Measurement Time30 s to 900 s
General Measurement FunctionsV/I waveform monitor, FFT analyzer, and inrush-current measurement
Standard InterfacesGPIB / RS232C / USB
Removable Data StorageCompactFlash card, up to 512 MB
AC Input100 V to 240 V AC, 50 Hz to 60 Hz
Dimensions430 W × 177 H × 270 D mm
Maximum Dimensions455 W × 195 H × 330 D mm
WeightApprox. 8 kg

Harmonic Current Standards

IEC 61000-3-2 Supports IEC 61000-3-2 Ed5.0 (2018), Ed3.0 (2005), and Ed2.2 (2004) for applicable single-phase equipment.
JIS C 61000-3-2 Supports JIS C 61000-3-2 (2019), (2005), and the applicable legacy test configuration.
IEC 61000-4-7 Measurement methods support IEC 61000-4-7 Ed2.1 (2009), Ed2.0 (2002), and Ed1.0 (1991).
Interharmonics Supports selectable processing according to the selected IEC 61000-4-7 measurement method.
Harmonic AnalysisThrough the 40th order in HA mode
Extended AnalysisThrough the 180th order in OTHER mode


Flicker & Voltage-Fluctuation Testing

IEC 61000-3-3 Supports IEC 61000-3-3 Ed3.1 (2017) and applicable earlier test configurations.
IEC 61000-4-15 Flicker measurement methods support IEC 61000-4-15 Ed2.0 (2010) and Ed1.1 (2003).
Pst Accuracy1 ±5%
Pst Measurement Time30 s to 900 s
Voltage-Fluctuation Measurement Selectable between simultaneous measurement with Pst and independent measurement.
Manual Switching dmax 3 to 24 measurements with 30 s to 180 s measurement time per test.


Standalone Compliance Testing

Test Condition SetupAvailable from the instrument front panel
Test ExecutionCan be performed without a connected PC
Real-Time JudgmentPass/fail results and spectrum information displayed during testing
Report CreationSupported directly from the KHA1000
Report FormatsPDF / text
Data StorageCompactFlash card, up to 512 MB
Optional PC SoftwareSD005-KHA Harmonics Explorer


Development & Diagnostic Measurements

Voltage MeasurementTrue RMS and peak
Current MeasurementTrue RMS and peak
Power MeasurementsActive, apparent, and reactive power
Power FactorSupported
Frequency45 Hz to 65 Hz
FFT AnalyzerBuilt-in
V/I Waveform MonitorBuilt-in
Maximum Direct Inrush80 Apeak within 20 ms
Repeatability Comparison Current test data can be compared with previously measured data to support repeatability evaluation.
Test Power Quality Check Measures voltage, frequency, peak voltage, and distortion of the AC test source before compliance testing.


Interfaces & Test System Integration

GPIBStandard
RS232CStandard
USBStandard communication interface
External AC Source ControlPCR-LE Series control through RS232C
AC Source Control ParametersVoltage, frequency, range, ON phase, and output ON/OFF
Application SoftwareSD005-KHA Harmonics Explorer
Line Impedance NetworkLIN Series available for applicable standards-based systems
Daily Pre-Test CheckerOP02-KHAS available