Home » Electronic Measuring Instruments » Power Supplies & Electronic Loads » DC Power Supply » KIKUSUI PWR401MH Compact Wide-Range DC Power Supply

KIKUSUI PWR401MH Compact Wide-Range DC Power Supply

KIKUSUI
Supplier / Manufacturer: KIKUSUI
Model: PWR401MH

$1,380

The KIKUSUI PWR401MH is a compact 400 W wide-range programmable DC power supply rated for up to 240 V and 5 A. It provides CV/CC operation for applications requiring higher DC voltage with moderate current in a compact 400 W platform.

LAN/LXI, USB, and RS232C are standard, with built-in sequence control, VMCB multi-channel operation, variable internal resistance, and remote sensing for automated test and R&D systems.

Industries & Applications:

240 V Wide-Range DC Power in a Compact 400 W Platform

The KIKUSUI PWR401MH is a compact programmable CV/CC DC power supply designed for applications that require output voltages up to 240 V while maintaining flexible voltage and current combinations within a 400 W output-power envelope.

The PWR401MH provides a rated output of 240 V, 5 A, and 400 W, with maximum settable values of 252 V and 5.25 A. Its wide-range architecture allows the user to select different voltage and current operating points without requiring separate fixed-range DC supplies, provided the total output remains within the unit's rated power capability.

240 V / 5 A Higher-voltage PWR-01 MH output range
400 W Compact wide-range output power
252 V / 5.25 A Maximum settable voltage and current
LAN / USB / RS232C Standard interfaces for automated control

Built-in hardware sequencing, analog control, variable internal resistance, preset memory, remote sensing, synchronized operation, and Virtual Multi-Channel Bus functionality make the PWR401MH suitable for both standalone engineering use and integrated automated test systems.

240 V / 5 A / 400 W Wide-Range Output

Provides higher-voltage DC output up to 240 V and current up to 5 A while supporting flexible operating points within the 400 W rated output-power limit.

Precision CV/CC Regulation

Constant-voltage setting accuracy is ±(0.05% of setting + 0.05% of rating), with ±26 mV line and load regulation. Constant-current setting accuracy is ±(0.5% of setting + 0.1% of rating).

Built-In 64-Step Sequence

One hardware sequence program with up to 64 steps can be stored in the power supply. Step time ranges from 0.1 seconds to 100 hours, and stored programs can run without a connected PC after programming.

Variable Internal Resistance

In CV mode, the PWR401MH can simulate source internal resistance from 0.01 Ω to 36.00 Ω for evaluating systems using battery, solar-cell, fuel-cell, or other source models.

LAN/LXI, USB, and RS232C

Standard digital interfaces support remote programming and automated testing. LAN supports LXI, VXI-11, SCPI-RAW, and HiSLIP communication.

VMCB Multi-Channel Control

Virtual Multi-Channel Bus technology allows one PC to control up to 31 PWR-01 units through a virtual multi-channel power-system architecture.

High-voltage DC electronic device and component evaluation
Programmable CV/CC laboratory power testing up to 240 V
Automated production and validation test systems
Multi-channel DC power systems using VMCB control
Battery and DC source simulation using variable internal resistance
Programmable voltage-profile and endurance testing using sequence control

PWR401MH Specifications

Operating Mode CV / CC
Rated Output Voltage 240 V
Rated Output Current 5 A
Rated Output Power 400 W
Maximum Settable Voltage 252 V
Maximum Settable Current 5.25 A
CV Setting Accuracy ±(0.05% of setting + 0.05% of rating)
CC Setting Accuracy ±(0.5% of setting + 0.1% of rating)
CV Line Regulation ±26 mV
CV Load Regulation ±26 mV
CC Line Regulation ±2.5 mA
CC Load Regulation ±6.0 mA
Transient Response 2 ms or less
CV Ripple Noise 100 mVp-p / 20 mVrms
CC Ripple Noise 12 mArms
Maximum Remote Sensing Compensation 5 V per line
Variable Internal Resistance 0.01 Ω to 36.00 Ω
Nominal AC Input 100 Vac to 240 Vac, 50 Hz to 60 Hz, single phase
AC Input Voltage Range 85 Vac to 265 Vac
Input Frequency Range 47 Hz to 63 Hz
Standard Interfaces LAN / LXI, USB, RS232C
Dimensions 71 W × 124 H × 350 D mm
Weight Approx. 3 kg

Programming & System Control

Hardware Sequence Programs 1 program
Hardware Sequence Steps Up to 64 steps
Sequence Step Time 0.1 s to 100 h
Sequence Repetition 1 to 99,998 or infinite
Interval Loops Up to 16 configurable interval loops
Synchronized Operation Voltage/current setting synchronization and sequence-step synchronization using trigger signals
Preset Memory Three sets of voltage, current, OVP, OCP, and UVL settings
VMCB Virtual multi-channel control for up to 31 PWR-01 units
Output ON/OFF Delay 0.0 s or 0.5 s to 99.9 s
Soft Start / Stop 0.0 s or 0.5 s to 10.0 s


Analog & Remote Control

External Voltage Programming 0% to 100% of rated output using selectable 0–5 V or 0–10 V control
External Current Programming 0% to 100% of rated output using selectable 0–5 V or 0–10 V control
External Resistance Control Supported for output voltage and current programming
Voltage Monitor Output Selectable 0–5 V or 0–10 V
Current Monitor Output Selectable 0–5 V or 0–10 V
External Output ON/OFF Supported
External Shutdown Supported
External Alarm Clear Supported
Communication Protocol SCPI / IEEE 488.2
LAN LXI, VXI-11, SCPI-RAW, HiSLIP


Expansion & Protection

Master-Slave Parallel Operation Up to three same-model 400 W units including the master
Maximum PWR401MH Parallel Current 15 A rated current with three units
Series Operation Up to two same-model units
Series One-Control Operation Not supported
Overvoltage Protection OVP
Overcurrent Protection OCP
Undervoltage Limit UVL
Additional Protection Overheat, incorrect sensing connection, low AC input, shutdown, power limit, communication watchdog, and master-slave parallel-operation protection