Home » Electronic Measuring Instruments » Flow & Airflow Measurement » Positive Displacement Flowmeter » OVAL FLOWPET-5G Positive Displacement Flowmeter

OVAL FLOWPET-5G Positive Displacement Flowmeter

OVAL FLOWPET-5G Positive Displacement Flowmeter

Measure fuel oil or water usage on boilers and industrial utility lines with on-site indication and selectable signal outputs.

The OVAL FLOWPET-5G is a positive displacement meter designed primarily for the oil and water supplied to boiler systems. A local electronic register displays instantaneous and cumulative flow, helping facilities track energy consumption and equipment operation.

Different measuring bodies and calibration specifications are provided for oil and water. Standard oil variants cover 7–24,000 L/h at 0–120°C, while high-temperature oil versions cover 14–6,400 L/h at temperatures up to 150°C. Water versions cover 200–12,000 L/h, with different accuracy and construction materials. These are overall series ranges rather than a universal operating span for one meter.

A dirt-resistant register places operational switches behind the screen. The front cover can be positioned to simplify access; battery pack replacement can be carried out without removing the meter from the pipeline. The display assembly angle can be adjusted over a 150° range to suit site conditions.

Output configurations include pulse, analog, alarm and combined analog-plus-pulse/alarm options. OVAL describes retrofit compatibility with earlier FLOWPET electronics, but compatibility, output assignment and replacement parts should be checked against the exact existing meter designation.

7–24,000 L/h Standard oil-series flow envelope
±0.5% reading Oil-version published accuracy
Up to 150°C High-temperature oil variant
150° adjustment Electronic register viewing angle

Positive Displacement Measurement

Measure volumetric oil or water throughput with a meter intended for plant utility and boiler monitoring.

Live and Accumulated Flow Readings

Read instantaneous flow and totalized consumption directly at the register.

Oil and Water Variants

Choose a matching wetted body, pressure rating, flow calibration and temperature range for the actual fluid.

Easier Register Handling

Rear-positioned switches reduce accidental touches; the register can be reoriented over a 150° range.

Serviceable Battery Pack

Replace the battery pack from the register without taking the meter body out of the installed pipe run.

Selectable Output Types

Integrate pulse, analog, alarm or combined signal arrangements with external management equipment as specified.

Boiler Fuel Monitoring

Measure kerosene, light oil or appropriate heavy fuel oil flow for boiler operations.

Boiler Feed Water

Track boiler-related water use with a dedicated water-service variant.

Plant Energy Accounting

Use totalized fuel or water quantities to assess consumption and operational efficiency.

Legacy Meter Modernization

Evaluate electronics replacement for compatible earlier FLOWPET assemblies without major piping modifications.

Industrial Utility Management

Transmit flow pulses or analog readings to production dashboards and energy-monitoring systems.

Routine Facility Maintenance

Support periodic visual checks, consumption logs and local operation reviews.

OVAL FLOWPET-5G Positive Displacement Flowmeter Specifications

Measurement principlePositive displacement volumetric measurement
Product designationFLOWPET-5G; OVAL family/type LS
Oil versionsStandard and high-temperature oil-service versions
Oil-service nominal sizes, standard20, 25, 40 and 50 mm
Oil-service nominal sizes, high temperature20, 25 and 40 mm
Water nominal sizes20, 25, 40 and 50 mm
Oil flow envelope, standard7–24,000 L/h across six types; manufacturer viscosity footnote applies
Oil flow envelope, high-temperature14–6,400 L/h across four types
Water flow envelope200–12,000 L/h across four types
Oil-service fluid typesKerosene, light oil and compatible heavy oil
Water-service fluidWater
Oil-service accuracyWithin ±0.5% of reading
Water-service accuracyWithin ±1% of reading
Operating temperature, standard oil0 to +120°C
Operating temperature, high-temperature oil0 to +150°C
Operating temperature, water0 to +120°C
Maximum pressure, standard oil1.18 MPa
Maximum pressure, high-temperature oil0.98 MPa
Maximum pressure, water1.18 MPa under manufacturer-specified static/running water conditions
Connections, oilJIS 10K flange or ASME Class 150 flange, model-dependent
Connections, waterJIS 10K flange
Body material, oil / waterFC250 cast iron / stainless steel
Power and signalsDC or battery; analog/pulse/alarm combinations depend on specification
Enclosure and area ratingIP65 under stated installation conditions; non-explosionproof

Model and Size Comparison

ConfigurationStandard oilHigh-temp. oilWater
Nominal diameter20–50 mm20–40 mm20–50 mm
Flow envelope7–24,000 L/h14–6,400 L/h200–12,000 L/h
Fluid temperature0 to 120°C0 to 150°C0 to 120°C
Max. pressure1.18 MPa0.98 MPa1.18 MPa*
Accuracy±0.5% rdg.±0.5% rdg.±1% rdg.

The published oil flow-range footnote refers to a viscosity basis of 2–5 mPa·s; actual measurement ranges differ across fuels and meter sizes. *Water pressure includes manufacturer-specific service conditions. IP65 is listed with an “install under the eaves” condition. This is not an explosionproof meter.