OVAL CoriMate II Coriolis Flowmeter

OVAL CoriMate II Compact Microflow Coriolis Flowmeter

Compact, single-tube Coriolis flowmeter for precise low-flow liquid measurement and small process lines.

The OVAL CoriMate II measures very small liquid mass flow rates in laboratory and light industrial applications. Its compact, single-tube design is available with integrated or remote mounting and does not require a separate vibration-isolated table under the manufacturer’s stated installation conditions.

The three nominal sensor bores—0.7, 1.5 and 3 mm—have respective maximum flow rates of 75, 300 and 2,700 g/min. Accuracy depends on the operating flow: below or at 33% of full-scale flow, the published value is ±0.1% of full scale (F.S.); above 33% and up to 100%, it is ±0.3% of reading (RD).

CoriMate II uses a DC power supply, provides analog and pulse output, and has an IP66 non-explosion-proof configuration. The manufacturer’s specification table limits the applicable process fluid to liquids, the operating temperature to −10 to +60°C and pressure to 2 MPa.

0.7 / 1.5 / 3 mmThree nominal flow-path sizes
2,700 g/minLargest model’s maximum liquid flow
2 MPaMaximum operating pressure
IP66Non-explosion-proof construction

Low-flow Coriolis measurement

Direct mass-flow metering with three bore sizes for microflow applications.

Single-tube design

Compact tube path simplifies handling and integration into small setups.

Multiple mounting layouts

Available as an integral or separately mounted instrument configuration.

No special isolation table required

Manufacturer states vibration-isolated tables are unnecessary for supported installations.

DC operation

Designed for DC power with analog and pulse measurement outputs.

Defined ingress protection

IP66, non-explosion-proof configuration for suitable environments.

Laboratory Liquid Microflow

Evaluate small liquid flow rates in benchtop research setups.

Precision Liquid Dosing

Monitor minute dosing and injection process streams.

Small-Scale Process Development

Compare changing liquid flow conditions on pilot equipment.

Chemical Liquid Feed Monitoring

Check compatible liquid feeds in compact transfer lines.

Liquid Equipment Characterization

Use repeatable mass-flow readings in instrument and process R&D.

Pulse-Output Monitoring

Connect pulse output to suitable data acquisition or counting systems.

SpecificationDetails
Nominal flow-path diameter0.7 mm | 1.5 mm | 3 mm
Maximum flow by diameter75 g/min | 300 g/min | 2,700 g/min
Applicable fluidLiquid
Accuracy, at or below 33% of full-scale flow±0.1% of full scale (F.S.)
Accuracy, over 33% to 100% of flow±0.3% of reading (RD)
Process connectionRc1/8
Operating temperature−10 to +60°C (non-condensing)
Maximum operating pressure2 MPa
Wetted materialSUS316L
Power supplyDC powered
OutputsAnalog and pulse
EnclosureIP66; non-explosion-proof
Mounting choicesIntegral or remote-mounted configuration

Do not label CoriMate II a gas or steam meter based on older narrative marketing copy: the manufacturer’s published applicable-fluid specification says liquid. The ±0.1% metric is a percentage of full scale for low flow, not ±0.1% of reading. Non-explosion-proof construction is not suitable where hazardous-area approvals are required.