Home » Optical / Imaging / Precision Devices » Light, Color & UV Measurement » Illuminance Meter » TechnoOptis IM-1000R Illuminance Spectrometer

TechnoOptis IM-1000R Illuminance Spectrometer

TechnoOptis IM-1000R Illuminance Spectrometer

Legacy handheld spectrometric illuminance measurement for lighting quality, chromaticity, color rendering and PPFD assessment.

The TechnoOptis IM-1000R is a legacy illuminance spectrometer that the manufacturer explicitly lists as discontinued. Its published specifications remain useful for identifying existing installations, replacement requirements and measurement compatibility.

It uses an LVF (linear variable filter) and silicon photodiode array to measure across 380–780 nm with 1 nm output wavelength intervals. The instrument supports illuminance (2–1,000,000 lx), spectral graphs, chromaticity, correlated color temperature, CRI (Ra and R1–R15) and photosynthetic photon flux density (PPFD).

The IM-1000R connects through RS-232C. It must not be presented as a wireless LAN model: the IM-1000 includes the wireless module, while the IM-1000R does not, as documented in the manufacturer manual.

The instrument is included here as a technical reference and for installed-base inquiries, not as a new-production offering. Contact Denkei Americas to identify available replacements, service support and appropriate current instruments for a specified wavelength and measurement application.

Discontinued Manufacturer-confirmed product status
380–780 nm Published spectral measurement range
2–1M lx Published illuminance measurement span
RS-232C IM-1000R interface; no built-in WLAN

Visible-Light Spectral Measurement

Use the published 380–780 nm sensor capability to analyze light sources.

Illuminance and Chromaticity

Review lux, xy, u′v′ and XYZ measurements from installed equipment.

CRI and Color Temperature

Evaluate Ra, R1–R15 and correlated color temperature on supported samples.

PPFD for Horticulture

Assess photosynthetically active photon flux density for grow-light studies.

Handheld Historical Configuration

Document the backlit display and battery-powered field measurement setup.

RS-232C Data Interface

Access existing instruments through the specified wired serial connection.

LED Lighting Development

Reference spectral and color rendering data for past LED qualification programs.

Architectural Lighting Surveys

Maintain traceability of legacy lux and CCT measurement records.

Roadway and Exterior Lighting

Understand measurement scope for historical public-lighting projects.

Automotive Lighting Evaluation

Compare spectral illuminance results from existing instrument fleets.

Horticultural Lighting

Evaluate PPFD records from plant-factory or grow-light installations.

Instrument Replacement Planning

Define the CRI, spectral, lux and communications requirements for successor selection.

IM-1000R Specifications

ProductIM-1000R illuminance spectrometer
AvailabilityDiscontinued by manufacturer
Measurement principleLVF (linear variable filter) plus silicon photodiode array
Measured wavelength span380–780 nm
Output wavelength interval1 nm (not necessarily optical resolution)
Illuminance measurement range2–1,000,000 lx; standard source A, AUTO range
Illuminance accuracy±2% ±1 digit under stated manufacturer conditions
Chromaticity accuracy at ≥50 lxxy ±0.0020
Chromaticity accuracy at 10–50 lxxy ±0.0035
Chromaticity accuracy at 5–10 lxxy ±0.0050
Illuminance repeatability±0.5% +1 digit under specified conditions
Spectral response deviation f1′2% or less
Cosine response f23% or less
Measurement outputsEv, XYZ, xy, u′v′, CCT, dominant wavelength, CRI Ra/R1–R15, spectral graph, PPFD
Measurement modeAUTO (full/first/adjust) or MANUAL
Manual acquisition exampleApprox. 0.2 s with STR2 command under specified conditions
Auto acquisition timeApprox. 0.5–50 s depending on range and setup
CommunicationRS-232C, wired; no onboard WLAN on IM-1000R
Display128 × 64 dot backlit LCD
PowerFour NiMH AA batteries or optional dedicated AC adapter
Battery runtimeApprox. 7 h with new fully charged batteries
Operating conditions-10 to +40°C; ≤85% RH noncondensing
Dimensions70 × 250 × 78 mm (W × D × H), excluding protrusions
WeightApprox. 640 g including battery

Discontinued product, for historical technical reference only. IM-1000R does not contain the IM-1000 wireless LAN module. No direct successor has been verified; contact Denkei Americas for a current configuration review.