TechnoOptis UVR-300 UV Radiometer

TechnoOptis
Supplier / Manufacturer: TechnoOptis
Model: UVR-300

Contact Us

High-Sensitivity UV Radiometer

U.S.-based sales, and delivery available. The UVR-300 is a high-sensitivity digital UV radiometer designed for industrial UV curing, photolithography, sterilization, and advanced R&D applications. Featuring a wide measurement range from 0.1 to 280,000 μW/cm², it supports multiple detector options covering 220–490 nm to meet diverse process requirements. With USB communication, analog output, auto/manual range switching, and deviation measurement functions, the UVR-300 enables precise UV intensity monitoring, process control, and performance evaluation across manufacturing and laboratory environments.

■ Wide range detection sensitivity : 0.1 – 280,000μW/cm2 ■ Multiple measurement function. ■ Capable of data communication to PC with USB and application software.

■ Industrial field of using the photochemical reaction ■ Production line with photolithography of SEMI, FPD, PCB and etc. ■ Evaluation and management of sensitive material for print industry ■ Performance evaluation of solar ■ Industrial field of using the photo electricity reaction ■ Electrophotographic print ■ Electrophotographic photoreceptor ■ Biochemistry field of using the biological reaction ■ Erythema reaction、Pigmentation evaluation ■ Skin photosensitivity reaction ■ Management of rearing environment for cattle farmer and fish farmer ■ Management of spindly growth and photosynthetic response ■ Others field of using the sterilizing activity ■ Food processing field ■ Medical cosmetic treatment field ■ Various R&D

ItemSpecification
UD-250UD-360AUD-400
Photo detectorSilicon photo diode
Measurement Range0.1 – 280,000μW/cm2
Measurement ModeAuto / Manual (4 range)
Wavelength Range220 – 300nm320 – 400nm360 – 490nm
Linearity±5% of rdg. ±1 digit (Auto Range)
Measurement functionStandard measurement:Radiant intensity (μW/cm2), Irradiation intensity (mJ/ cm2) Deviation measurement:Intensity difference (μW/cm2), Intensity ratio (%)
Analog Output0~3V、 1mV/1digit
InterfaceUSB
Power SourceSize AA dry-cell battery × 2 or AC adapter (Option)
Light source for calibrationGL-15(*1)FL20S.BLB(*2)Standard illuminant A
Temperature drift : fTwithin ±3% ( -10 – +40℃ with 23℃ as reference)
Humidity drift : fHwithin ±3%
Use conditionsTemperature : -10 – +40℃, Humidity : 85% RH or less (No condensation)
Outer dimensions (W×D×H)195×70×33mm (with detector unit)
Weight260g (with battery weight)
(*1) Sterilization lamp (*2) Black light