TechnoOptis SR-5AS Spectroradiometer

TechnoOptis SR-5AS Spectroradiometer

Flagship HDR spectral luminance analysis with a 0.0001 cd/m² low-end specification for ultra-dark display performance.

The TechnoOptis SR-5AS is the flagship SR-5-series spectroradiometer for ultra-low-luminance and extreme HDR optical measurements. Its cooled linear CCD captures visible spectral radiance from 380 to 780 nm, allowing engineers to evaluate luminance, CIE color coordinates, XYZ and correlated color temperature.

At a 2° measuring angle the SR-5AS has a published lower limit of 0.0001 cd/m², making it suitable for very dark OLED and HDR display states. At 0.1° the published upper limit is 500,000,000 cd/m². These values describe different measurement angles and must not be interpreted as one identical range for every aperture.

With motorized 2°, 1°, 0.2° and 0.1° fields of view, touchscreen operation, high-speed measurement and external synchronizing functions, the SR-5AS supports both laboratory characterization and automated optical evaluation.

The SR-5AS extends sensitivity relative to the SR-5A at the 2° field of view. Select the actual model and aperture against the light source, required chromaticity limits and integration setup before ordering.

0.0001 cd/m² Ultra-low luminance at 2°
500 M cd/m² High-end limit at 0.1°
380–780 nm Visible spectral radiance
4 angles Motorized measuring fields

0.0001 cd/m² Low-Luminance Capability

Examine deep blacks and ultra-dim light at the specified 2° measurement angle.

Extreme High-Luminance Coverage

Capture high-brightness emission up to 500 million cd/m² at 0.1°.

Cooled Linear CCD Sensor

Acquire wavelength-resolved radiance data via a diffraction-grating optical system.

Four Motorized Measurement Angles

Switch measurement aperture for small device regions and different brightness regimes.

Synchronized High-Speed Measurement

Support optical analysis of flickering or pulsed sources using external vertical sync and high-speed modes.

PC Integration and Touch Control

Operate locally using the touchscreen or use supported PC connectivity for automated collection.

OLED Ultra-Black Evaluation

Measure near-black states and dark-scene emission on OLED panels.

Micro LED Development

Characterize small emitters across very dim and bright operating states.

HDR Display Qualification

Compare panel luminance and color behavior over a large dynamic span.

Automotive Instrument Panels

Measure color and luminance of cockpit displays and indicators.

High-Brightness LED Testing

Evaluate intense LED and specialty illumination with an appropriate field angle.

Spectral Gamma and IVL Studies

Support wavelength-resolved luminance and color analysis in R&D or production setups.

SR-5AS Specifications

ModelSR-5AS
DetectorElectronically cooled linear CCD
Dispersive elementDiffraction grating
Objective / viewfinderf = 82 mm, F2.5 objective; 5° viewfinder field
Measuring angles2° / 1° / 0.2° / 0.1°; motorized selection
Spectral measurement range380–780 nm
Wavelength resolution1 nm
Spectral wavelength accuracy±0.3 nm, measured with mercury emission lamp
Spectral measurement quantitySpectral radiance (W·sr⁻¹·m⁻²·nm⁻¹)
Photometric/colorimetric measurementsRadiance, luminance, tristimulus XYZ, CIE 1931 xy, CIE 1976 u′v′, correlated color temperature/duv, and spectral peak; 2°/10° standard observer
Reference luminance accuracy±2% or less, under manufacturer reference conditions
Polarization sensitivityLuminance ≤1%; spectral radiance ≤2% over 400–780 nm
SupplyDedicated AC adapter; 100–240 V AC, 50/60 Hz; 12 V DC instrument input
WeightApprox. 5.5 kg
Measuring distance250 mm to infinity, from objective-lens metal tip
Spectral bandwidth5 nm FWHM
Luminance range — 2°0.0001–1,500,000 cd/m²
Luminance range — 1°0.0005–4,500,000 cd/m²
Luminance range — 0.2°0.0125–100,000,000 cd/m²
Luminance range — 0.1°0.05–500,000,000 cd/m²
Chromaticity accuracy (x, y)x ±0.0015; y ±0.001 above manufacturer-stated luminance thresholds
Chromaticity accuracy threshold — 2°Applies from 0.0005 cd/m², not the 0.0001 cd/m² detectability threshold
Luminance repeatability1.5%: 0.0005–0.005 cd/m²; 0.4%: 0.005–0.1 cd/m²; 0.3%: ≥0.1 cd/m²
Chromaticity repeatability0.005: 0.0005–0.005 cd/m²; 0.0015: 0.005–0.1 cd/m²; 0.0005: ≥0.1 cd/m²
Measurement / external syncAuto/manual integration; high-speed/synchronous operation and vertical-sync input
Digital interfacesRS-232C and USB 3.0
Power consumptionApprox. 30 W
Operating conditions5 to 30°C; ≤80% RH, noncondensing
Dimensions (W × D × H)422 × 130 × 244 mm

Measurement area by field angle

Angle250 mm350 mm500 mm1,000 mm
2°6.5 mm10.0 mm15.1 mm32.2 mm
1°3.25 mm4.99 mm7.55 mm16.1 mm
0.2°0.65 mm1.00 mm1.51 mm3.22 mm
0.1°0.33 mm0.50 mm0.76 mm1.61 mm

Measurement distances are from the objective-lens metal fixture tip; spot diameters shown in the table are manufacturer reference data. Ranges/accuracy are conditional on aperture, light source and test conditions.

The 0.0001 cd/m² value is a measurement limit at 2°, not the lower threshold for the quoted chromaticity accuracy; the manufacturer specifies chromaticity performance only above higher angle-dependent luminance thresholds.

Measuring-angle upper and lower luminance values are based on standard illuminant A and specified modes; spectral accuracy is referenced to a Hg emission lamp.

Accessories such as attachment lenses, neutral-density filters and an illuminance adapter are not automatically included.