Home » Optical / Imaging / Precision Devices » Light, Color & UV Measurement » Illuminance Meter » TechnoOptis IM-600 / IM-600M Digital Illuminance Meter

TechnoOptis IM-600 / IM-600M Digital Illuminance Meter

TechnoOptis IM-600 / IM-600M Digital Illuminance Meter

JIS AA-class digital lux measurement with a wide range and a choice of standard or compact sensing-head configurations.

The TechnoOptis IM-600 and IM-600M are digital illuminance meters for lighting measurement, product qualification and building illumination surveys. Both models measure from 0.005 to 999,000 lux with automatic or manual six-step ranging.

IM-600 uses a standard sensing design; its display and photodetector can be separated using optional extension cables. The IM-600M employs a compact 16 × 21 mm photodetector head, better suited to integration where mounting space is limited.

The two versions differ in cosine correction performance, particularly at high angles of incidence. At 80°, the manufacturer lists deviation within ±10% for IM-600 and within ±25% for IM-600M; head size should therefore be balanced against the angular response requirement.

Supported functions include instantaneous illuminance, integrated illuminance, exposure/integration time, luminous intensity, and deviation measurements. Built-in USB connectivity and 0–3 V analog output support compatible external data collection and oscilloscopes. A dedicated AC adapter and extension cable options are separate accessories.

0.005–999k lx Published illuminance measurement range
JIS AA Referenced JIS C 1609-1:2006 class
16 × 21 mm IM-600M compact photodetector head
USB / 0–3 V Digital and analog outputs

Wide Lux Measurement Range

Measure low to high illumination levels over six ranges.

JIS AA-Class Design

Apply a manufacturer-specified standard for professional illuminance testing.

Two Detector Configurations

Choose IM-600 standard sensor or IM-600M compact head.

Integrated-Light Functions

Report illuminance, lux-hours, measurement duration and ratios.

USB and Analog Data Paths

Connect to supported PCs or monitor voltage output for lighting response.

Color Correction Settings

Use the color correction factor setting for appropriate light-source workflows.

Lighting Fixture Development

Characterize lux output of LED, OLED and fluorescent luminaires.

Building Lighting Audits

Check task illumination in commercial and industrial interiors.

Factory and Warehouse Maintenance

Inspect work area light levels and plan upkeep.

Plant-Growth Lighting

Track illuminance as part of lighting management, distinct from PPFD.

Production-Line Sensor Integration

Install a compact sensing head in confined equipment.

Pulsed Lighting Observation

Monitor brightness response of compatible PWM-driven lighting with an oscilloscope.

IM-600 / IM-600M Specifications

ProductsIM-600 and IM-600M digital illuminance meters
Instrument classGeneral AA class per JIS C 1609-1:2006
DetectorSilicon photodiode
Illuminance measurement range0.005–999,000 lx
RangesSix-step auto/manual
Linearity±2% of reading ±1 digit, auto range
Relative spectral response error≤6% per manufacturer conditions
IM-600 cosine correction at 80°Within ±10%
IM-600M cosine correction at 80°Within ±25%
IM-600 cosine correction at 60°Within ±5%
IM-600M cosine correction at 60°Within ±7%
Standard measurementsIlluminance (lx), integrated illuminance (lx·h), integration time (h), luminous intensity (cd)
Deviation measurementsIlluminance difference (lx), illuminance ratio (%)
Color correctionConfigurable CCF (color correction factor)
Analog output0–3 V
Digital interfaceUSB
PowerTwo AA dry-cell batteries; dedicated AC adapter optional
Main unit dimensions188 × 65 × 34 mm (W × D × H)
IM-600M receiving-head size16 × 21 mm
Approx. weight without batteriesIM-600: 200 g; IM-600M: 230 g
Operating environment-10 to +40°C; ≤85% RH noncondensing
Optional accessoryExtension cables for compatible IM-600 arrangements

Angular cosine correction by variant

Incident angleIM-600IM-600M
10°Within ±1%Within ±1%
30°Within ±1%Within ±2%
50°Within ±5%Within ±6%
60°Within ±5%Within ±7%
80°Within ±10%Within ±25%

Published specifications are angle-, range-, mode- and configuration-dependent; accessories, approvals, calibration and availability should be verified for the selected order code.