HIOKI SS7081-50 Battery Cell Voltage Generator

12-Channel Battery Cell Simulation for BMS Validation

The HIOKI SS7081-50 generates and measures simulated battery cell voltages across 12 independent channels to validate battery management systems without using live battery cell strings.

Each channel combines programmable DC voltage output with source/sink current and measurement capability. This enables BMS engineers to reproduce cell voltage conditions, current balance behavior and abnormal states such as cell shorts or open-wire conditions.

12 Channels Independent cell-voltage simulation
5.025 V Maximum programmable DC output per channel
±1 A Maximum source/sink current per channel
1,000 V Maximum permitted series configuration voltage

Voltages are programmable independently from 0 to 5.0250 V, and instrument series connections can be configured for high-voltage BMS tests up to a maximum combined output of 1,000 V. Low-current measurements use a dedicated 100 µA range for BMS standby current and leakage evaluation.

LAN control supports automated test setups and software-driven sequences. HIOKI specifies that the control PC, control software and BMS wiring are not included with the instrument; clarify the required software download, cables and test-system design before procurement.

Twelve Independent Cell Channels

Configure and measure twelve simulated cell voltages within a single chassis.

Precision DC Voltage Generation

Program independent outputs from 0.0000 to 5.0250 V with stated output accuracy of ±0.0150% of setting ±500 µV.

Two-Quadrant Current Control

Sources or sinks up to 1 A per channel for cell balancing and BMS test scenarios.

Fault-Condition Simulation

Simulates cell shorts and open-wire conditions without constructing a string of live battery cells.

Low-Current Measurement

Includes a 100 µA current range for leakage and low-power BMS circuit measurement.

Scalable Automated Test Systems

Supports instrument series configurations up to 1,000 V and LAN-driven test sequences.

Battery Management System Validation

Check voltage sensing and control response under defined cell states.

Cell Balancing Circuit Testing

Source and sink current to evaluate active or passive balancing response.

Open-Wire Fault Simulation

Exercise BMS error detection for disconnected cell monitoring lines.

Cell Short-Circuit Simulation

Assess BMS diagnostic behavior under controlled abnormal-cell scenarios.

BMS Dark Current Measurement

Monitor standby and leakage current using the low-current measurement range.

High-Voltage Pack Simulator Development

Configure multiple instruments within the 1,000 V series limit for large battery module BMS tests.

Specification SS7081-50
Independent channels 12
Programmable DC voltage, per channel 0.0000 to 5.0250 V
Maximum source/sink current, per channel ±1.00000 A
Maximum instrument series output 1,000 V (configuration limit)
DC voltage measurement range −0.00100 to 5.10000 V
Current measurement ranges ±1.20000 A or ±120.0000 µA
Voltage output accuracy ±0.0150% of setting ±500 µV
Voltage measurement accuracy ±0.0100% rdg. ±100 µV
Current measurement accuracy, 1 A range ±0.0700% rdg. ±100 µA
Current measurement accuracy, 100 µA range ±0.0350% rdg. ±10 nA
Interface LAN
Power input 100–240 V AC, 50/60 Hz
Dimensions / mass 430 × 132 × 483 mm; approx. 10.3 kg
Standard accessories User manual, power cord, rack frame
Required separately Control PC, control software and BMS wiring

The 1,000 V figure is a maximum permitted series configuration, not the output voltage of a single 12-channel unit. Control PC, software and BMS wiring are not standard supplied items; confirm system requirements and safe high-voltage test design.