Combined AC Withstand Voltage and Insulation Resistance Testing
The HIOKI 3174 AC Automatic Insulation/Withstanding HiTESTER combines up to 5 kV AC dielectric withstand testing with 500 V and 1,000 V DC insulation resistance testing, contact verification and programmable test sequences.
Rated at 100 VA for AC withstand testing, the 3174 uses PWM-controlled output to reduce sensitivity to AC supply fluctuations. It measures leakage current from 0.01 mA to 20.0 mA and supports selectable 50/60 Hz test frequency.
Contact-check functionality helps identify improper test-lead connections; a second 9615 high-voltage test-lead set must be purchased separately for contact checks. Digital window comparators support pass/fail inspection.
Remote operation is available with HIOKI Safety Test Data Management Software 9267, while programmable ramp and delay intervals help standardize production inspection. Verify high-voltage safeguards and fixture requirements before integration.