HIOKI RM3545A Precision Resistance Meter

Ultra-High-Precision DC Resistance Measurement for Weld and Connection Quality Inspection

The HIOKI RM3545A is a precision DC resistance meter designed for accurate measurement of extremely low resistance values in EV battery connections, motor windings, electronic components, and automated production inspection systems. With 1 nΩ minimum resolution and up to 1 A measurement current, it supports detailed evaluation of electrical connection quality.

The RM3545A provides 13 measurement ranges from 1000 µΩ to 1000 MΩ, covering low-resistance welds, precision shunt resistors, connector contacts, and higher-resistance electrical components. Its four-terminal measurement method reduces the influence of test lead and contact resistance for accurate measurements.

1 nΩ Minimum resistance resolution
13 Ranges 1000 µΩ to 1000 MΩ
1 A DC Maximum measurement current
20 Channels Optional internal multiplexing on RM3545A-2

The RM3545A series includes two configurations. The RM3545A-1 provides single-channel resistance measurement, while the RM3545A-2 supports installation of up to two optional Z3003 multiplexer units for a maximum of 20 four-terminal measurement channels or 42 two-terminal channels.

For larger multi-point inspection systems, the RM3545A supports compatible external HIOKI switching solutions. With an SW1001 switch mainframe, configurations can support up to 33 four-terminal channels, while an SW1002 configuration can support up to 132 four-terminal channels.

Advanced measurement capabilities include Offset Voltage Compensation (OVC), Pure Resistance (PR) mode, temperature correction, contact checking, low-power measurement, and comparator-based PASS/FAIL judgment. These functions help manufacturers identify resistance variations associated with welding defects, connection inconsistencies, and component quality.

Standard LAN, USB, RS-232C, and EXT I/O interfaces enable integration with automated testing systems. Measurement speeds can be as fast as 2.3 ms under specified conditions, while precision low-resistance testing in PR mode supports high-speed measurements from 21 ms.

The RM3545A includes a Z2001 temperature sensor. Measurement leads must be selected separately for the standard instrument configurations. The built-in Z3003 multiplexer is supported only on the RM3545A-2 model.

1 nΩ Ultra-High Resolution

Measures extremely low DC resistance with 1 nΩ resolution in the 1000 µΩ range, supporting precise inspection of welds, electrical connections, and low-resistance components.

13 Measurement Ranges

Covers resistance ranges from 1000 µΩ to 1000 MΩ with up to 1 A measurement current and high-accuracy four-terminal measurement.

Optional Multi-Channel Measurement

The RM3545A-2 supports up to two Z3003 multiplexer units for 20 four-terminal or 42 two-terminal channels. Compatible external switch mainframes support larger systems.

High-Speed Production Inspection

Provides measurement times as fast as 2.3 ms in supported settings and 21 ms in specified low-resistance Pure Resistance (PR) mode conditions.

Advanced Resistance Measurement Functions

Includes offset voltage compensation, temperature correction, contact checking, low-power measurement, and comparator functions to improve inspection reliability.

LAN and Automated Test Integration

Provides standard Ethernet, USB, RS-232C, EXT I/O, and analog output for integration with automated inspection and production testing systems.

EV Battery Busbar Weld Inspection

Evaluate resistance variations across welded battery connections to support detection of insufficient welds and electrical connection defects.

Motor and Transformer Winding Testing

Measure DC winding resistance of motors, transformers, and coils during development, production, and quality inspection.

Multi-Point Electrical Inspection

Combine the RM3545A-2 with optional Z3003 multiplexers or compatible external switching systems to inspect multiple electrical connections.

Connector and Contact Resistance Testing

Evaluate electrical resistance in connectors, relays, switches, terminals, and wiring assemblies using appropriate measurement conditions.

Shunt Resistor and Component Testing

Verify resistance values of precision shunts, fuses, and other components used in electronic and power conversion systems.

Automated Production Quality Control

Use remote control, comparator judgments, and high-speed resistance measurements in manufacturing inspection and quality assurance systems.

Specification RM3545A-1 RM3545A-2
Model Configuration Single-channel model Internal multiplexer-compatible model
Resistance Ranges 13 ranges, 1000 µΩ to 1000 MΩ 13 ranges, 1000 µΩ to 1000 MΩ
Minimum Resolution 1 nΩ 1 nΩ
Maximum Measurement Current 1 A DC 1 A DC
Internal Multiplexer Support Not supported Up to two Z3003 units
Maximum Internal 4-Wire Channels 1 20 with two Z3003 units
Maximum Internal 2-Wire Channels Single-channel configuration 42 with two Z3003 units
Internal Multiplexer Switching Time Not applicable 30 ms
External Switch Mainframe Supported Supported
Dimensions 215 W × 80 H × 306.5 D mm 215 W × 80 H × 306.5 D mm
Mass Approx. 2.7 kg Approx. 3.4 kg

The RM3545A-2 requires optional Z3003 multiplexer units for multi-channel operation. The RM3545A-1 does not support internal Z3003 installation. External switching configurations and maximum channel counts depend on the selected switch mainframe and installed modules.


Specification RM3545A Series
Measurement Method DC four-terminal method
Representative Accuracy, 1000 µΩ Range ±(0.045% rdg. + 0.010% f.s.)
Representative Accuracy, 10 mΩ Range ±(0.045% rdg. + 0.001% f.s.)
Representative Accuracy, 100 mΩ Range ±(0.045% rdg. + 0.001% f.s.)
Representative Accuracy, 1000 mΩ Range ±(0.012% rdg. + 0.001% f.s.)
Representative Accuracy, 1000 Ω Range ±(0.006% rdg. + 0.001% f.s.)
Measurement Speed, FAST 2.3 ms (representative, range-dependent)
Measurement Speed, MED 22 ms at 50 Hz / 19 ms at 60 Hz
Measurement Speed, SLOW1 102 ms
Measurement Speed, SLOW2 202 ms
PR Mode, 10 mΩ Range, FAST 21 ms
Maximum Open-Terminal Voltage 8 V for ranges up to 1000 Ω; 20 V for 10 kΩ and above
Low-Power Measurement Supported; maximum open-terminal voltage 20 mV
Temperature Measurement -10.0°C to 99.9°C with Z2001 sensor
Offset Voltage Compensation OVC supported
Temperature Correction TC and temperature conversion (ΔT)
Contact Check Supported
Comparator Supported
LAN 10BASE-T / 100BASE-TX, TCP/IP
USB Supported
RS-232C Up to 115200 bps
EXT I/O D-sub 37-pin
Analog Output DC 0 to 1.5 V
Power Supply AC 100 to 240 V, 50/60 Hz
Maximum Power Consumption 40 VA
Accuracy Guarantee Period 1 year
Included Temperature Sensor Z2001
Measurement Leads Sold separately with standard main-unit configurations

Representative accuracy specifications are based on HIGH mode, OVC enabled, SLOW2, and no zero adjustment. Accuracy varies by range and measurement settings. Measurement times are representative and depend on the selected resistance range and operating mode. The 21 ms PR-mode value applies to specified low-resistance measurement conditions and external triggering.

Internal Z3003 multiplexing is available only with RM3545A-2. External switch mainframe configurations can support up to 33 four-terminal channels using SW1001 or 132 four-terminal channels using SW1002, depending on module configuration. Confirm the permissible voltage and wiring configuration of the complete measurement system. Standard measurement probes are sold separately.

Resistance Range Maximum Display Resolution Measurement Current
1000 µΩ1200.000 µΩ1 nΩ1 A
10 mΩ12.00000 mΩ10 nΩ1 A
100 mΩ120.0000 mΩ100 nΩ1 A
1000 mΩ1200.000 mΩ1 µΩ100 mA
10 Ω12.00000 Ω10 µΩ10 mA
100 Ω120.0000 Ω100 µΩ10 mA
1000 Ω1200.000 Ω1 mΩ1 mA
10 kΩ12.00000 kΩ10 mΩ1 mA
100 kΩ120.0000 kΩ100 mΩ100 µA
1000 kΩ1200.000 kΩ1 Ω10 µA
10 MΩ12.00000 MΩ10 Ω1 µA
100 MΩ120.0000 MΩ100 Ω100 nA (high-precision mode)
1000 MΩ1200.0 MΩ100 kΩ1 µA or less

Measurement currents are based on HIOKI's published range specifications. Actual measurement current and available functions vary by measurement mode. Minimum resolution is not the same as measurement accuracy.