HIOKI CT6700 Current Probe

Low-Noise 50 MHz Current Waveform Measurement for Oscilloscopes

The HIOKI CT6700 is a high-sensitivity AC/DC current probe for observing current waveforms from DC to 50 MHz, including low-current transients in power electronics, switching supplies and control circuits.

Rated for 5 A rms continuous input with frequency-dependent derating, the CT6700 provides a 1 V/A output for viewing low-level current signals on compatible oscilloscopes and recorders. Its magnetic shielding helps reduce interference from nearby electric fields.

DC–50 MHz Wide measurement frequency bandwidth
5 A rms Continuous allowable current, with derating
1 V/A Probe output conversion ratio
≤7 ns Specified rise time

The probe offers a typical noise level of 60 μA rms, supporting observation of current waveforms on the order of 1 mA. One-button demagnetization and zero adjustment help restore measurement stability after high-current tests.

A compatible probe power supply, such as the HIOKI 3269, is required. Use a measuring instrument with at least 1 MΩ input impedance and follow HIOKI precautions when connecting metallic BNC terminals to recorders with plastic BNC sockets.

DC to 50 MHz Bandwidth

Observes fast switching and transient currents from DC through high-frequency spectral content.

High Sensitivity for Small Currents

A low-noise output and 1 V/A conversion support analysis of milliamp-level current waveforms.

5 A Continuous Input Rating

Measures up to 5 A rms under defined conditions, subject to frequency derating.

Fast Transient Response

Rise time of 7.0 ns or shorter supports observation of rapid current transitions.

Automatic Demagnetization and Zeroing

One-button operation helps correct offsets and residual magnetization before measurement.

Oscilloscope and Recorder Integration

BNC waveform output operates with high-impedance instruments and an appropriate dedicated probe power supply.

Switching Power Supply Analysis

Observe turn-on, turn-off and ripple-related current waveforms in power converters.

Inverter and Motor-Drive Development

Inspect switching transients and control currents in power-electronics equipment.

Low-Current Waveform Measurement

Characterize milliamp-level dynamic current behavior with high signal-to-noise performance.

Digital Control Circuit Evaluation

Monitor fast current changes in control and drive circuits.

Oscilloscope-Based R&D

Add non-invasive current waveform measurement to high-impedance oscilloscope channels.

Transient Response Troubleshooting

Investigate short-duration current events that are difficult to observe with slower instruments.

Specification Details
Frequency Bandwidth DC to 50 MHz (−3 dB)
Rise Time 7.0 ns or shorter
Continuous Allowable Input 5 A rms (requires derating with frequency)
Maximum Non-Continuous Peak Input ±7.5 A peak
Output Ratio 1 V/A; internally terminated
Noise Level 60 μA rms typical; 75 μA rms maximum (30 MHz-band instrument)
Amplitude Accuracy, Typical ±1% rdg. ±1 mV (DC/45–66 Hz; 0–5 A rms)
Amplitude Accuracy, Guaranteed ±3% rdg. ±1 mV (DC/45–66 Hz; 0–5 A rms)
Maximum Conductor Diameter φ 5 mm; insulated conductors
Probe Power Supply ±12 V ±0.5 V, 3.2 VA
Recommended Power Supply HIOKI 3269 (separate)
Required Instrument Input High input impedance of 1 MΩ or greater
Sensor Dimensions 155 W × 18 H × 26 D mm
Terminator Dimensions 29 W × 83 H × 40 D mm
Mass Approx. 250 g
Sensor Cable Length 1.5 m to BNC terminal
Power Cable Length 1 m

Use only on insulated conductors; do not exceed current and frequency derating limits. A compatible power supply is required. HIOKI warns that metallic BNC connectors can damage some plastic recorder BNC terminals.