Home » Electronic Measuring Instruments » Power Supplies & Electronic Loads » Source Measure Unit » ADCMT 6240B DC Voltage Current Source/Monitor

ADCMT 6240B DC Voltage Current Source/Monitor

Bipolar Source Measure Unit for Pulsed Device and Low-Resistance Testing

The ADCMT 6240B combines high-speed pulsed voltage and current sourcing with 5½-digit measurement and precision low-resistance testing in one bipolar source measure unit.

The 6240B provides DC voltage output up to ±15 V at ±1 A and pulsed current up to ±4 A at ±10 V. Its 50 μs minimum pulse width supports device characterization while reducing measurement errors caused by DUT self-heating.

±15 V / ±1 A DC bipolar source and sink operation
±4 A Pulse Source or sink for pulses up to 20 ms
50 μs Minimum pulse width with 1 μs resolution
0.25 μΩ Minimum resistance measurement resolution

Four sweep types—fixed, linear, random, and two-slope linear—support semiconductor and electronic component characterization. Separate HI and LO limiter settings help protect batteries, capacitors, laser diodes, and other devices that are sensitive to over-discharge or reverse voltage.

The integrated low-resistance measurement function can cancel thermal electromotive force generated at metal contact surfaces, supporting precise testing of connectors, conductive materials, wire harnesses, and shunt resistors.

Bipolar Source and Sink Operation

Source or sink positive and negative voltage and current, including 0 V sink operation for battery and power-device evaluation.

High-Current Pulse Output

Generate or sink pulse current up to ±4 A for a maximum pulse width of 20 ms and duty cycle of 20% or less.

Flexible Source and Sweep Modes

Select DC, pulse, DC sweep, or pulse sweep operation with fixed, linear, random, and two-slope linear sweep types.

Independent HI/LO Limiters

Configure HI and LO limiters independently to control charging and discharging limits and protect voltage-sensitive devices.

Thermal EMF Cancellation

Reverse current polarity and cancel thermal electromotive force for precision contact-resistance and low-resistance measurement.

Suspend Function

Select STBY, HiZ, or LoZ output-off states to reduce relay operation, improve throughput, and limit transient current when connecting batteries.

Semiconductor and electronic component characterization
Diode and LED forward-voltage testing
Battery charge and discharge testing
Power-device and power-IC evaluation
Connector contact-resistance measurement
Wire-harness resistance testing
Conductive material and shunt-resistor measurement
Automated production-line characteristic testing

6240B Specifications

Source Resolution 4½ digits
Output Method Bipolar source and sink
Maximum DC Output ±15 V / ±1 A
Maximum Pulse Output ±10 V / ±4 A
Maximum ±4 A Pulse Width 20 ms at 20% duty cycle or less
Minimum Source Resolution — Voltage 10 μV
Minimum Source Resolution — Current 1 nA
Measurement Resolution 5½ digits, maximum ±319,999 counts
Basic Source and Measurement Accuracy ±0.02%, representative range
Minimum Measurement Resolution — Voltage 1 μV
Minimum Measurement Resolution — Current 100 pA
Resistance Measurement Up to 0.75 GΩ
Minimum Resistance Resolution 0.25 μΩ
Minimum Pulse Width 50 μs with 1 μs setting resolution
Minimum Pulse Cycle 2 ms with measurement
500 μs without measurement
Source Modes DC, pulse, DC sweep, pulse sweep
Sweep Types Fixed, linear, random, two-slope linear
Interfaces USB, GPIB