KYOWA PCD-400A Sensor Interface

Four-Channel USB Interface for Strain Gages and Transducers

The KYOWA PCD-400A is a compact, PC-connected carrier-type sensor interface for acquiring strain-gage and strain-gage-transducer signals in laboratory and engineering test setups.

Four input channels are available per instrument, with synchronized measurement expandable to 16 channels by stacking up to four PCD-400A units. A USB-connected PC can configure the measurement and collect data using compatible KYOWA acquisition software.

4 channelsStrain measurement channels per interface
16 channelsMaximum with four synchronized instruments
10 kHzMaximum sampling frequency; not analog bandwidth
AC 2 VrmsSpecified strain-bridge excitation

The interface supports quarter-, half- and full-bridge strain-gage connections using compatible input adapters. Its specified measurement-response range is DC to 200 Hz, distinct from the maximum 10 kHz digital sampling frequency.

The PCD-400A appears in KYOWA’s discontinued-product catalog archive. For a new system, ask Denkei about the current PCD-400B Series and verify adapters, software, power and mechanical compatibility before treating it as a direct replacement.

Four Channels in One Interface

Acquire four bridge-type strain measurement channels from a compact sensor interface.

Synchronized Multi-Unit Operation

Use up to four connected interfaces for 16-channel synchronized measurement.

PC-Based USB Acquisition

Configure and collect measurements from a compatible PC rather than relying on an onboard screen.

Flexible Bridge Connections

Support quarter-, half- and full-bridge gage connections with suitable wiring and adapters.

Multiple Strain Ranges

Select among seven nominal strain ranges from 200 to 20,000 microstrain full scale.

Digital Acquisition and Filtering

Up to 10 kHz sampling, with selectable low-pass filters, for strain-signal evaluation.

Component Strain Testing

Monitor strain on mechanical parts during controlled loading and bench experiments.

Transducer Evaluation

Read compatible strain-gage-based force, load or pressure transducers.

Prototype Development

Capture short-duration strain signals during engineering design verification.

Multi-Point Stress Studies

Synchronized measurements across up to 16 channels in a stacked interface configuration.

Material Test Fixtures

Connect multiple gage locations to a PC for laboratory loading tests.

Legacy Test-System Support

Document existing PCD-400A installations and plan modernization of KYOWA measurement systems.

SpecificationDetails
Measuring targetsStrain gages and strain-gage transducers
Input channels4 per PCD-400A
Multi-unit synchronizationUp to 4 instruments / 16 channels
Input formBalanced differential; D-sub 37-pin connector
Bridge systemsQuarter bridge 120 Ω (2-/3-wire); half/full bridge 120–1000 Ω
Bridge excitationAC 2 Vrms
Gage factor2.00 fixed
Strain ranges200, 500, 1,000, 2,000, 5,000, 10,000, 20,000 µε
Nominal range accuracyWithin ±0.5% full scale; consult model calibration terms
Response frequency rangeDC to 200 Hz (within ±10% response)
Maximum sampling frequency10 kHz; not equivalent to 10 kHz analog bandwidth
Analog-to-digital conversion24-bit
InterfaceUSB connection to compatible PC/software
Product statusPCD-400A is listed in manufacturer discontinued-product catalog archive

Specifications refer to legacy PCD-400A, not PCD-400B/430B. Compatible adapters, power source, PC software and replacement availability must be verified for the installation.