HIOKI BT3563A Battery HiTESTER

High-Voltage Battery Internal Resistance and Voltage Testing up to 300 V

The HIOKI BT3563A Battery HiTESTER provides high-speed, high-precision measurement of battery internal resistance and open-circuit voltage for automated production-line inspection of large lithium-ion battery packs and modules up to 300 V DC.

Using a 1 kHz AC four-terminal measurement method, the BT3563A measures internal resistance across seven ranges from 3 mΩ to 3 kΩ, with minimum resolution of 0.1 µΩ. Its 6 V, 60 V, and 300 V voltage ranges support inspection of individual cells, battery modules, and high-voltage battery assemblies within the instrument's rated limits.

±300 V DC Maximum rated battery input voltage
0.1 µΩ Minimum resistance resolution
1 kHz AC-IR Four-terminal resistance measurement
LAN Standard Ethernet interface for automated inspection

The BT3563A simultaneously measures AC internal resistance (AC-IR) and open-circuit voltage (OCV), helping manufacturers evaluate battery quality during production, shipping, and acceptance inspections. Independent resistance and voltage comparators provide Hi, IN, and Lo judgments for automated inspection decisions.

Fast measurement modes support high-throughput testing, while standard LAN, RS-232C, and 37-pin EXT I/O interfaces enable integration with PLC-controlled equipment and production-line automation systems. Additional functions include contact checking, zero adjustment, averaging, statistical calculations, and analog output.

Compared with the previous BT3563 generation, the BT3563A offers improved tolerance to measurement lead resistance, standard LAN connectivity, and an electrostatic-discharge-resistant design tested at ±30 kV under specified IEC 61000-4-2 conditions. These enhancements support reliable operation and reduced maintenance requirements in automated test environments.

Measurement leads are sold separately. For high-voltage battery testing, select compatible probes, cables, and switching equipment rated for the complete measurement system.

Simultaneous AC-IR and OCV Measurement

Measures battery internal resistance using a 1 kHz AC four-terminal method together with open-circuit voltage for production-line quality evaluation.

High-Voltage Measurement up to 300 V DC

Supports battery modules and large EV battery packs with 6 V, 60 V, and 300 V measurement ranges and basic voltage accuracy of ±0.01% of reading ±3 digits.

0.1 µΩ Resistance Resolution

Provides seven resistance measurement ranges from 3 mΩ to 3 kΩ, with 0.1 µΩ minimum resolution for low-resistance battery cells and electrical connections.

High-Speed Production-Line Inspection

Supports fast sampling modes, independent resistance and voltage comparators, contact checking, and external trigger functions for automated battery inspection.

Standard LAN and PLC Connectivity

Includes 10BASE-T/100BASE-TX LAN, RS-232C, 37-pin EXT I/O, and analog output interfaces for integration into automated test equipment.

Improved Lead Resistance and ESD Tolerance

Offers enhanced tolerance to measurement lead resistance and an input design tested against ±30 kV contact electrostatic discharge under IEC 61000-4-2 conditions.

EV Battery Pack Production Testing

Measure internal resistance and open-circuit voltage of high-voltage battery packs up to 300 V during automated end-of-line inspection.

Battery Module Quality Control

Evaluate electrical characteristics of assembled lithium-ion battery modules during production and quality verification.

Battery Shipping and Acceptance Inspection

Compare internal resistance and voltage measurements against defined acceptance criteria before shipment or integration into larger battery systems.

Automated Battery Inspection Systems

Integrate resistance and voltage measurements with PLC-controlled equipment, automated fixtures, and production-line sorting processes.

Battery Tab and Busbar Evaluation

Use high-resolution low-resistance measurement to investigate suitable welded connections, tabs, and conductive interconnects.

Battery Research and Reliability Testing

Record resistance and open-circuit voltage data across battery samples, production batches, and controlled evaluation conditions.

Specification BT3563A
Measurement Functions AC Internal Resistance (AC-IR) and DC Open-Circuit Voltage (OCV)
Resistance Measurement Method AC four-terminal method
Resistance Measurement Frequency 1 kHz ±0.2 Hz
Resistance Measurement Ranges 3 mΩ / 30 mΩ / 300 mΩ / 3 Ω / 30 Ω / 300 Ω / 3 kΩ
Minimum Resistance Resolution 0.1 µΩ (3 mΩ range)
Resistance Basic Accuracy ±0.5% rdg. ±10 dgt. (3 mΩ range); ±0.5% rdg. ±5 dgt. (30 mΩ and higher ranges)
Rated Input Voltage ±300 V DC
Maximum Rated Voltage to Earth ±300 V DC
Voltage Measurement Ranges 6 V / 60 V / 300 V DC
Minimum Voltage Resolution 10 µV (6 V range)
Voltage Basic Accuracy ±0.01% rdg. ±3 dgt.
Response Time 10 ms (specified response time; separate from sampling period)
Sampling Period, Ω + V, 60 Hz 8 ms (EX.FAST) / 24 ms (FAST) / 70 ms (MEDIUM) / 253 ms (SLOW)
Sampling Period, Ω or V, 60 Hz 4 ms (EX.FAST) / 12 ms (FAST) / 35 ms (MEDIUM) / 150 ms (SLOW)
Sampling Period, Ω + V, 50 Hz 8 ms (EX.FAST) / 24 ms (FAST) / 84 ms (MEDIUM) / 259 ms (SLOW)
Sampling Period, Ω or V, 50 Hz 4 ms (EX.FAST) / 12 ms (FAST) / 42 ms (MEDIUM) / 157 ms (SLOW)

Resistance Range Maximum Display Resolution Measurement Current
3 mΩ3.1000 mΩ0.1 µΩ100 mA
30 mΩ31.000 mΩ1 µΩ100 mA
300 mΩ310.00 mΩ10 µΩ10 mA
3 Ω3.1000 Ω100 µΩ1 mA
30 Ω31.000 Ω1 mΩ100 µA
300 Ω310.00 Ω10 mΩ10 µA
3 kΩ3.1000 kΩ100 mΩ10 µA

Voltage Range Maximum Display Resolution
6 V 6.00000 V 10 µV
60 V 60.0000 V 100 µV
300 V 300.000 V 1 mV
Specification BT3563A
ModelBT3563A
Measurement ModesResistance + Voltage, Resistance Only, Voltage Only
ComparatorIndependent resistance and voltage judgments: Hi / IN / Lo
Contact CheckSupported
Zero Adjustment±1000 counts
Statistical CalculationsUp to 30,000 measurements
Additional FunctionsPulse measurement, delay, averaging, panel save/load, memory storage
LAN InterfaceTCP/IP, 10BASE-T / 100BASE-TX (standard)
RS-232CUp to 38.4 kbps; printer interface supported
External I/O37-pin handler interface
Analog OutputDC 0–3.1 V
PC Software SupportLabVIEW driver available
Electrostatic Discharge Resistance±30 kV contact discharge under specified IEC 61000-4-2 test conditions
Power Supply100–240 V AC, 50/60 Hz
Power Consumption35 VA maximum
Dimensions215 W × 80 H × 295 D mm
MassApprox. 2.4 kg
Accuracy Guarantee Period1 year
Included AccessoriesInstruction manual, power cord, operating precautions
Measurement LeadsSold separately

Specifications are based on HIOKI's published BT3563A product information. The ±300 V DC rating applies to the instrument's rated input and voltage to earth; it does not authorize use above the ratings of connected probes, fixtures, cables, or switching systems. Voltage resolution depends on the selected range. Measurement accuracy and sampling time depend on operating conditions. Response time and sampling period are separate specifications.

For high-voltage battery measurements, select suitable measurement leads, such as HIOKI L2100 or L2110, according to the application. HIOKI SW1001/SW1002 switch mainframes are rated for 60 V DC maximum input and must not be assumed suitable for switching 300 V battery measurements.