HIOKI BT3562A Battery HiTESTER

High-Speed Battery Internal Resistance and Voltage Testing up to 100 V

The HIOKI BT3562A Battery HiTESTER is designed for high-speed, automated inspection of lithium-ion battery cells, modules, and mid-sized battery packs. It simultaneously measures AC internal resistance (AC-IR) and open-circuit voltage (OCV) for battery quality control and production testing up to 100 V DC.

The BT3562A uses a 1 kHz AC four-terminal resistance measurement method to evaluate battery internal resistance from the 3 mΩ range through 3 kΩ. Its minimum resistance resolution of 0.1 µΩ supports detailed evaluation of low-resistance battery cells, electrode connections, and conductive interconnects.

100 V DC Maximum rated battery voltage
0.1 µΩ Minimum internal resistance resolution
1 kHz AC-IR Four-terminal resistance measurement
LAN Standard Ethernet connectivity for automated systems

The instrument provides seven resistance ranges (3 mΩ to 3 kΩ) and three voltage ranges (6 V, 60 V, and 100 V). High-speed sampling, independent resistance and voltage comparator functions, and external I/O support repeatable PASS/FAIL inspections in battery manufacturing environments.

Standard LAN, RS-232C, and a 37-pin EXT I/O handler interface enable integration with PLC-controlled production equipment. The BT3562A also provides analog output and measurement functions including contact checks, zero adjustment, averaging, and statistical calculations.

Compared with the earlier BT3562, the BT3562A extends the maximum rated measurement voltage from 60 V to 100 V and adds standard LAN connectivity. Improved tolerance to measurement lead resistance and an electrostatic-discharge-resistant design help simplify automated test-system development and maintenance.

Measurement leads are sold separately. Select appropriate probes and fixtures according to the voltage rating, battery terminal geometry, and production-line configuration.

Simultaneous AC-IR and OCV Measurement

Measures battery internal resistance using a 1 kHz AC four-terminal method alongside DC open-circuit voltage for production-line quality assessment.

High-Resolution Low-Resistance Testing

Offers seven resistance ranges from 3 mΩ to 3 kΩ, with minimum resolution of 0.1 µΩ for low-resistance battery cells and electrical connections.

100 V Battery Voltage Measurement

Provides 6 V, 60 V, and 100 V DC measurement ranges with basic voltage accuracy of ±0.01% of reading ±3 digits.

High-Speed Automated Inspection

Supports fast measurement modes, resistance and voltage comparators, contact checking, and external trigger/control functions for automated inspection systems.

Standard LAN and PLC Integration

Includes 10BASE-T/100BASE-TX Ethernet, RS-232C, and a 37-pin EXT I/O handler interface for production-line automation and remote control.

Improved Noise and ESD Resistance

Provides improved measurement lead resistance tolerance and a design rated to withstand ±30 kV contact electrostatic discharge under IEC 61000-4-2 test conditions.

Lithium-Ion Battery Cell Inspection

Measure internal resistance and open-circuit voltage during cell production, final inspection, and quality verification.

EV Battery Module Testing

Evaluate electrical characteristics of automotive battery modules and mid-sized battery packs within the 100 V measurement rating.

Battery Shipping and Acceptance Inspection

Compare battery resistance and voltage against defined acceptance criteria before shipment or assembly into larger systems.

Automated Battery Production Lines

Integrate high-speed resistance and voltage measurements into PLC-controlled inspection equipment and automated sorting stations.

Battery Tab and Busbar Evaluation

Use high-resolution low-resistance measurements to investigate suitable welded connections, tabs, and conductive interconnects.

Battery Research and Quality Validation

Compare resistance and open-circuit voltage across battery samples, production batches, and validation conditions.

Specification BT3562A
Measurement Functions AC Internal Resistance (AC-IR) and DC Open-Circuit Voltage (OCV)
Resistance Measurement Method AC four-terminal method
Resistance Measurement Frequency 1 kHz ±0.2 Hz
Resistance Ranges 3 mΩ / 30 mΩ / 300 mΩ / 3 Ω / 30 Ω / 300 Ω / 3 kΩ
Minimum Resistance Resolution 0.1 µΩ (3 mΩ range)
Resistance Basic Accuracy ±0.5% rdg. ±10 dgt. (3 mΩ range); ±0.5% rdg. ±5 dgt. (30 mΩ and higher ranges)
Maximum Rated Input Voltage ±100 V DC
Maximum Rated Voltage to Earth ±100 V DC
Voltage Measurement Ranges 6 V / 60 V / 100 V DC
Minimum Voltage Resolution 10 µV (6 V range)
Voltage Basic Accuracy ±0.01% rdg. ±3 dgt.
Response Time 10 ms (specified response time; separate from sampling period)
Sampling Period, Ω + V, 60 Hz 8 ms (EX.FAST) / 24 ms (FAST) / 70 ms (MEDIUM) / 253 ms (SLOW)
Sampling Period, Ω or V, 60 Hz 4 ms (EX.FAST) / 12 ms (FAST) / 35 ms (MEDIUM) / 150 ms (SLOW)

Resistance Measurement Ranges

Range Maximum Display Resolution Measurement Current
3 mΩ3.1000 mΩ0.1 µΩ100 mA
30 mΩ31.000 mΩ1 µΩ100 mA
300 mΩ310.00 mΩ10 µΩ10 mA
3 Ω3.1000 Ω100 µΩ1 mA
30 Ω31.000 Ω1 mΩ100 µA
300 Ω310.00 Ω10 mΩ10 µA
3 kΩ3.1000 kΩ100 mΩ10 µA

Voltage Measurement Ranges

Range Maximum Display Resolution
6 V6.00000 V10 µV
60 V60.0000 V100 µV
100 V100.000 V1 mV
Specification BT3562A
ModelBT3562A
ComparatorResistance and voltage judgment: Hi / IN / Lo
Contact CheckSupported
Zero Adjustment±1000 counts
Statistical CalculationsUp to 30,000 measurements
Additional FunctionsPulse measurement, delay, averaging, panel save/load, memory storage
LANTCP/IP, 10BASE-T / 100BASE-TX (standard)
RS-232CUp to 38.4 kbps; printer interface supported
External I/O37-pin handler interface
Analog OutputDC 0–3.1 V
PC Software IntegrationLabVIEW driver available
Electrostatic Discharge Resistance±30 kV contact discharge, IEC 61000-4-2 test conditions
Power Supply100–240 V AC, 50/60 Hz
Power Consumption35 VA maximum
Dimensions215 W × 80 H × 295 D mm
MassApprox. 2.4 kg
Accuracy Guarantee Period1 year
Included AccessoriesInstruction manual, power cord, operating precautions
Measurement LeadsSold separately

Specifications are based on HIOKI's published BT3562A product data. Resistance and voltage accuracy depend on range and measurement conditions. Response time and sampling period are separate specifications. Simultaneous resistance and voltage sampling values shown above are based on 60 Hz mains settings; 50 Hz settings differ. Measurement leads are not included and must be selected according to the application and voltage rating.

For multi-channel testing, confirm the voltage rating of the complete switching system. HIOKI SW1001 and SW1002 switch mainframes are rated for a maximum of 60 V DC input and should not be assumed suitable for 100 V measurement setups.