HIOKI 3153 Automatic Insulation / Withstanding HiTESTER

Programmable AC/DC Hipot and Insulation Resistance Testing

The HIOKI 3153 Automatic Insulation / Withstanding HiTESTER is an electrical safety instrument combining AC and DC dielectric withstand testing with programmable DC insulation resistance measurement. It supports repeatable automated test sequences for production inspection and component validation.

Withstanding-voltage tests operate over 0.2–5.00 kV in AC or DC mode, while insulation test voltage is programmable from 50 V to 1,200 V DC. The instrument offers 32 program files, each with up to 50 steps, and control of voltage ramp-up and ramp-down timing.

5 kV Maximum AC/DC withstand voltage
50–1,200 V Programmable DC insulation test
32 files Stored test programs
50 steps Maximum steps per program

The tester uses PWM output control to stabilize the applied test voltage and provides voltage, current and resistance monitoring. The optional HIOKI 3930 High Voltage Scanner can support suitable multipoint test configurations.

For projects requiring AC withstand testing without DC hipot, compare the HIOKI 3174 AC Automatic Insulation/Withstanding HiTESTER.
The 3159 / 3159-02 series is discontinued and is documented separately for legacy replacement searches.

Programmable AC/DC Withstand Testing

Perform withstand testing from 0.2 to 5.00 kV in either AC or DC mode.

Programmable Insulation Voltage

Select 50–1,200 V DC for insulation resistance testing.

Multi-Step Automatic Programs

Create up to 32 files with up to 50 steps in each test sequence.

PWM Test-Voltage Control

Generate stable test voltage with reduced dependency on supply fluctuations.

Test-Voltage Ramp Control

Configure ramp-up and ramp-down timing as part of withstand test conditions.

Optional Multi-Point Testing

Expand suitable test setups using HIOKI’s compatible high-voltage scanner.

Production Electrical Safety Testing

Run repeatable safety inspection sequences on qualified electrical products.

Insulation Resistance Validation

Evaluate insulation resistance at configurable DC test voltages.

AC Dielectric Withstand Tests

Apply programmed AC hipot tests to compatible devices and components.

DC Dielectric Withstand Tests

Apply DC hipot procedures where supported by the test standard.

Automated Quality Control

Store and repeat programmed steps for consistent factory inspection.

Multi-Point Insulation Inspection

Add compatible scanner equipment to reduce manual reconnection.

Specification HIOKI 3153
AC withstand voltage 0.2–5.00 kV AC
DC withstand voltage 0.2–5.00 kV DC
AC withstand capacity 500 VA (maximum 30 minutes)
DC withstand capacity 50 VA (continuous)
Withstand voltage setting resolution 0.01 kV
AC test frequency 50 / 60 Hz selectable
Withstand current measurement 0.01–100.0 mA
Insulation test voltage 50–1,200 V DC (1 V steps)
Insulation measurement range 0.10–9,999 MΩ
Stored test programs 32 files
Maximum steps per file 50
Withstand test time 0.3–999 s
Voltage ramp / delay 0.1–99.9 s
Power supply 100–120 V or 200–240 V AC, 50/60 Hz
Dimensions (W × H × D) 320 × 155 × 480 mm
Mass 18 kg

The 500 VA AC output is rated for a maximum of 30 minutes; DC withstand capacity is 50 VA continuous. Use the actual equipment manual and relevant product test standards for safe installation and test configuration.