Home » Optical / Imaging / Precision Devices » Precision Measurement Equipment » Vibration Meter » ONO SOKKI LV-1800 Laser Doppler Vibrometer

ONO SOKKI LV-1800 Laser Doppler Vibrometer

Non-contact vibration velocity measurement from 0.3 Hz to 3 MHz

The ONO SOKKI LV-1800 Laser Doppler Vibrometer uses optical heterodyne detection to measure vibration velocity without attaching a contact accelerometer to the target.

It supports measurements from ultrasonic equipment and piezoelectric devices through MEMS and small electronic assemblies where sensor mass or physical contact would alter the vibration being studied. Standard bandwidth is 0.3 Hz to 3 MHz, and the maximum detected velocity is 10 m/s (zero-to-peak).

A variable-focus optical head supports stand-off distances from approximately 100 mm to 10 m. The positioning camera LV-0181 is an optional module, not an included standard camera. The LV-0800 optional low-velocity range board provides a minimum velocity resolution of 0.05 µm/s or less under its specified configuration; the standard 0.01 (m/s)/V range is 0.3 µm/s or less.

For frequency-domain measurement and engineering diagnostics, connect the analog velocity output to a suitable spectrum or FFT analyzer. Measurement results depend on the selected range, target reflectivity, stand-off distance, optical setup and optional equipment.

0.3 Hz–3 MHz Standard vibration frequency range
10 m/s Maximum velocity, zero-to-peak
100 mm–10 m Standard focus distance range
0.05 µm/s Optional LV-0800 velocity resolution or better

Optical, Non-contact Detection

Measure motion without adding an accelerometer or other mass to delicate or high-frequency test pieces.

Wide 3 MHz Measurement Band

Analyze vibration from 0.3 Hz through 3 MHz using the standard measurement ranges.

High Velocity Dynamic Capability

Measure vibration velocity up to 10 m/s zero-to-peak when the suitable instrument range is selected.

Optional Fine Velocity Resolution

Add the LV-0800 board for a low-velocity measurement range and specified resolution down to 0.05 µm/s or less.

Optional Visual Positioning

Use the LV-0181 camera module to observe the target and align the laser spot on a connected Windows PC.

Analog Output for Analysis

Send the demodulated vibration velocity waveform to compatible FFT or data-acquisition instrumentation.

Ultrasonic Tool Characterization

Investigate welding tools, ultrasonic cutters, cleaners and other high-frequency mechanical oscillators.

MEMS Device Vibration Testing

Characterize micromechanical resonators and shock sensors without loading them with contact probes.

Piezoelectric Component Development

Evaluate motion of piezo elements, ceramic capacitors and precision actuators.

EV Inverter Component Analysis

Study vibration of motor-related components and switching-device assemblies in development environments.

Electronics and Optical Device R&D

Measure micro-vibration in optical pickups, speakers, thin films and small electronic components.

Laboratory Vibration Spectroscopy

Feed non-contact velocity signals to an FFT analyzer for frequency response or modal investigations.

ONO SOKKI LV-1800 Laser Doppler Vibrometer Specifications

Measurement principleOptical heterodyne detection / vibration velocity demodulation
Standard frequency range0.3 Hz to 3 MHz (−3 dB)
Optional low-velocity range bandwidth0.3 Hz to 200 kHz in 0.001 (m/s)/V range
Maximum velocity10 m/s zero-to-peak (20 m/s peak-to-peak)
Standard minimum velocity resolution0.3 µm/s or less at 0.01 (m/s)/V
Optional minimum velocity resolution0.05 µm/s or less with LV-0800 low-velocity range board
Standard velocity output ranges1.0 / 0.1 / 0.01 (m/s)/V; optional 0.001 (m/s)/V
Velocity analog output±10 V; BNC; output impedance 50 Ω
Optical focus distanceApprox. 100 mm to 10 m (infinity indicated)
Minimum spot sizeApprox. 20 µm or less at 100 mm focus; smaller with optional microscope optics
Optional positioning cameraLV-0181, USB 2.0, VGA (640 × 480) at 30 fps
Signal cable length3 m (sensor and converter system)
Converter dimensions410 × 120 × 324 mm (W × H × D, excluding projections)
Approximate system mass8.1 kg including sensor cable
Operating temperature0 to +40°C
Operating humidity30–80% RH, noncondensing
Power input100–240 V AC, 50/60 Hz
Power consumptionApprox. 60 VA
CoolingNatural air cooling

The LV-0181 camera, LV-0800 board and microscope objectives are optional. Do not describe the 0.05 µm/s resolution as a standard-system value. Observe the applicable laser safety information in the selected configuration.