MedTEST Medical Electrical Safety Compliance System
Configurable IEC/UL 60601-1 safety testing for medical-device development, validation and end-of-line inspection.
The Keisoku Giken MedTEST brings together multiple medical electrical-safety test functions in a configured system rather than a single universal stand-alone instrument. It is intended to help manufacturers establish repeatable test sequences for equipment with Type B, BF and CF applied-part requirements under standards such as IEC/UL 60601-1.
The standard example configuration combines an 8207 Safety Compliance Analyzer with an SC6540 matrix scanner. The system can perform AC/DC dielectric withstand, insulation resistance, protective earth continuity, touch/patient leakage-current and energized RUN checks, with automatic selection among connected test points to reduce repeated lead changes.
The 8207 system example has an internal isolated AC source rated 500 VA and DUT testing voltage up to 277 V. The manufacturer describes a standard 16 A DUT-power test configuration, whereas up to 40 A mains handling requires expansion using the 620L touch-current instrument and suitable external power architecture, including 8500-series equipment. The 40 A earth-continuity test current is a separate function and must not be confused with 40 A DUT power handling.
Test-specific measuring-device (MD) networks, single-fault conditions, scanner capacity, connected DUT power and required cable/fixture options depend on the selected system. MedTEST supports engineering validation workflows; suitability for a regulatory approval program must be assessed against the applicable standard edition and certified test procedures.