Home » Electronic Measuring Instruments » Electrical Safety & Insulation Testing » Electrical Safety Tester » Keisoku Giken MedTEST Medical Device Safety Standards Compliance System

Keisoku Giken MedTEST Medical Device Safety Standards Compliance System

MedTEST Medical Electrical Safety Compliance System

Configurable IEC/UL 60601-1 safety testing for medical-device development, validation and end-of-line inspection.

The Keisoku Giken MedTEST brings together multiple medical electrical-safety test functions in a configured system rather than a single universal stand-alone instrument. It is intended to help manufacturers establish repeatable test sequences for equipment with Type B, BF and CF applied-part requirements under standards such as IEC/UL 60601-1.

The standard example configuration combines an 8207 Safety Compliance Analyzer with an SC6540 matrix scanner. The system can perform AC/DC dielectric withstand, insulation resistance, protective earth continuity, touch/patient leakage-current and energized RUN checks, with automatic selection among connected test points to reduce repeated lead changes.

The 8207 system example has an internal isolated AC source rated 500 VA and DUT testing voltage up to 277 V. The manufacturer describes a standard 16 A DUT-power test configuration, whereas up to 40 A mains handling requires expansion using the 620L touch-current instrument and suitable external power architecture, including 8500-series equipment. The 40 A earth-continuity test current is a separate function and must not be confused with 40 A DUT power handling.

Test-specific measuring-device (MD) networks, single-fault conditions, scanner capacity, connected DUT power and required cable/fixture options depend on the selected system. MedTEST supports engineering validation workflows; suitability for a regulatory approval program must be assessed against the applicable standard edition and certified test procedures.

IEC 60601-1Medical electrical safety workflow
AC 5 kV / DC 6 kVWithstand test ratings for 8207
8 / 16 pointsSC6540 switch configurations
500 VAIsolated source in standard setup

Multi-Function Medical Safety Testing

Combine AC/DC hipot, insulation resistance, ground bond, touch current and RUN tests in one project workflow.

Test-Point Switching with SC6540

Switch multiple points through an 8- or 16-channel matrix configuration without repeatedly moving test leads.

Applied-Part and Leakage Tests

Configure measuring-device networks and touch/patient leakage tests for relevant Type B, BF or CF requirements.

Built-In 500 VA Source

The 8207-based standard system includes isolated 0–277 V AC DUT power for applicable powered tests.

Optional High-Power Expansion

Select 620L and an appropriate external AC power system for higher current and larger medical DUTs.

Single-Fault and Production Sequences

Coordinate polarity/neutral/ground conditions and automated PASS/FAIL workflows as required by the test plan.

Medical Equipment Production QA

Run repeatable insulation, leakage-current and withstand checks at manufacturing test stations.

IEC 60601-1 Design Verification

Evaluate medical device electrical safety under prescribed normal and single-fault conditions.

Patient-Applied-Part Testing

Select suitable measuring networks and leakage current methods for B, BF and CF parts.

Multi-Point Device Testing

Use an SC6540 scanner to step through multiple leads, connectors and conductive parts.

Powered Medical Device Inspections

Perform supported RUN and touch-current measurements while powering the DUT with the selected source.

Safety Test-System Integration

Combine analyzer, scanner, fixtures, power sources and software into an automated inspection cell.

MedTEST System Configuration and Electrical Specifications

Specifications are configuration-dependent. The 8207/SC6540 standard system and expanded 620L/8500-based systems must not be presented as having identical DUT power or leakage-current ranges.

Main System Architecture

SpecificationStandard 8207 + SC6540Expanded 620L / 8500 Configuration
System conceptMedical electrical safety analyzer plus multipoint scannerAdditional touch-current/power instruments where required
AC withstand capabilityUp to 5 kV / 50 mAUp to 5 kV / 50 mA for the cited analyzer; accessory restrictions apply
DC withstand capabilityUp to 6 kV / 20 mAUp to 6 kV / 20 mA for the cited analyzer; accessory restrictions apply
Insulation resistance test voltageUp to 1,000 V DCUp to 1,000 V DC (analyzer-dependent)
Insulation resistance rangeUp to 50 GΩ (system published figure)Up to 50 GΩ for cited analyzer (confirm combination)
Protective earth / bondUp to 40 A earth-continuity testSame base test; not the DUT input-current rating
Touch/leakage-current functionsSupported; MD-network-dependentAdditional current ranges and scenarios with 620L
DUT supplyInternal isolated AC, 500 VAExternal isolated power system, roughly 500 VA–6 kVA examples
DUT AC supply voltage0–277 V (configuration-rated)0–277 V as described for full-expansion test setup
DUT mains current16 A class (standard 8207-based setup)Up to 40 A single-phase when appropriately expanded
Scanner test pointsSC6540 configurations with 8 or 16 channels8 or 16 channels per cited system configuration
Powered / RUN testSupportedSupported; requires appropriate power and safety architecture
IEC/UL 60601-1Test workflows relevant to medical equipmentTest workflows relevant to medical equipment

Additional Configuration Details

Subsystem / FunctionPublished Detail / Limitation
SC6540 selector statesH = high voltage; L = return/ground; X = open
Measuring-device networkIEC 60601-1 MD3 plus other MD circuit configurations; external MD option
Normal and single-fault scenariosPolarity, neutral and protective-ground switching per wiring/test setup
Built-in AC source limits0–277 V AC, 0–4.2 A and 500 VA; maximums are not simultaneously independent ratings
AC continuity / ground bondUp to 40 A, 8 V class and 600 mΩ cited for standard system
Touch current, base systemManufacturer cites 16 A/277 V DUT handling and 10 mA RMS/10 mA peak-to-peak class; refer to relevant MD mode
Touch current, expanded systemManufacturer cites 40 A/277 V DUT handling and 20 mA RMS/30 mA peak-to-peak class
Pre-operation check accessoriesTVB-2 (electrical safety tester) and LVB-2 (leakage-current tester)
Compliance assessmentSystem capability is not itself a certification or an assertion that every IEC 60601-1 test is supported in every configuration