KYOWA PCD-400B Sensor Interface

4-Channel USB Sensor Interface for Strain Gage Measurement

The KYOWA PCD-400B is a four-channel USB sensor interface designed for PC-based measurement of strain gages and strain-gage transducers. It combines bridge excitation, signal conditioning, and 24-bit A/D conversion in a compact instrument for laboratory testing, product evaluation, and automated strain data acquisition.

Each PCD-400B provides four balanced differential input channels for compatible quarter-bridge, half-bridge, and full-bridge strain measurement configurations. With the optional stacking kit, up to four units can be combined to create a synchronous measurement system with a maximum of 16 channels.

4 Channels Balanced differential strain inputs per unit
16 Channels Maximum synchronous stacked configuration
10 kHz Maximum sampling frequency
24-bit A/D conversion resolution

The PCD-400B supports seven strain measurement ranges from ±200 to ±20,000 microstrain, with a specified measurement accuracy within ±0.5% of full scale. Its AC 2 Vrms bridge excitation and automatic resistance and capacitance balance adjustment simplify configuration of compatible strain measurement circuits.

KYOWA specifies a frequency response from DC to 200 Hz and a maximum sampling frequency of 10 kHz. These are separate characteristics: the 10 kHz sampling specification does not mean that the instrument has a 10 kHz analog measurement bandwidth.

USB 2.0 connectivity enables operation with compatible PC software, including KYOWA DCS-100A dynamic data acquisition software. Optional input adapters support strain gages and strain-gage transducers, while selected adapters provide TEDS sensor information reading.

The PCD-400B is also compatible with KYOWA PCAS-100A software for printed circuit board strain and stress evaluation. For a complete PCB stress measurement system, KYOWA offers the PCAS-1000B set combining the PCD-400B, dedicated software, and applicable accessories.

The PCD-400B is designed for strain measurement. Applications requiring a dedicated voltage measurement mode should consider the PCD-430B. A compatible DC power source and appropriate input adapters must be selected for the intended system.

Four Balanced Differential Input Channels

Measures strain gages and compatible strain-gage transducers through four balanced differential channels with built-in bridge excitation.

Expandable to 16 Synchronous Channels

Supports stacking of up to four PCD-400B Series units using compatible accessories for synchronized multi-channel strain measurements.

24-Bit A/D Conversion

Uses 24-bit analog-to-digital conversion and seven selectable strain ranges for detailed recording of mechanical strain signals.

Up to 10 kHz Sampling

Supports sampling frequencies up to 10 kHz, including synchronous acquisition across 16 channels, with a specified strain measurement frequency response of DC to 200 Hz.

Automatic Bridge Balancing

Provides automatic resistance balance adjustment and capacitance self-tracking to simplify setup and compensate supported bridge measurement conditions.

USB and Software Integration

Connects to a PC through USB 2.0 and supports compatible KYOWA software for dynamic data acquisition, analysis, and PCB strain evaluation.

Printed Circuit Board Strain Testing

Measure strain during PCB assembly, handling, and mechanical testing to support reliability assessment and process improvement.

Mechanical Stress and Strain Analysis

Acquire strain gage measurements from components and structures subjected to static or changing mechanical loads.

Load Cell and Force Transducer Measurement

Record signals from compatible strain-gage-based load cells and force transducers using suitable input adapters.

Multi-Point Strain Measurement

Combine up to four sensor interface units for synchronized measurement at multiple locations across a test specimen.

Product Reliability and Durability Testing

Monitor mechanical strain during engineering validation, structural evaluation, and manufacturing quality investigations.

Research and Laboratory Data Acquisition

Collect strain measurement data through PC-based acquisition software for analysis, reporting, and experimental research.

Specification PCD-400B
Measurement TargetsStrain gages and strain-gage transducers
Input Channels4 per unit
Input ConfigurationBalanced differential inputs
Maximum Synchronous Channels16 channels using up to 4 stacked units
Input ConnectorD-sub 37-pin
Quarter-Bridge Compatibility120 Ω, 2-wire or 3-wire
Half-Bridge Compatibility120 to 1000 Ω
Full-Bridge Compatibility120 to 1000 Ω
Bridge Excitation2 Vrms AC
Gage Factor2.00 fixed
Strain Ranges200 / 500 / 1000 / 2000 / 5000 / 10000 / 20000 µε
Strain Range AccuracyWithin ±0.5% FS
NonlinearityWithin ±0.1% FS
Frequency ResponseDC to 200 Hz; deviation within ±10%
Maximum Sampling Frequency10 kHz
16-Channel SamplingUp to 10 kHz synchronous sampling with four units
A/D Converter Resolution24 bits
Low-Pass FilterSecond-order Butterworth; 10 Hz / 30 Hz / 100 Hz / FLAT
Resistance Balance AdjustmentAuto balance, within ±2%
Capacitance Balance AdjustmentCST automatic tracking, within 5000 pF
TEDS Sensor SupportInformation reading with UI-10A or UI-11A input adapters
Measurement Setting StorageNonvolatile memory

Specification PCD-400B
PC InterfaceUSB 2.0 High-Speed; compatible with USB 3.0 ports
Power Supply11 to 16 VDC
Power ConnectorHirose RM12BRD-4PH
Current Consumption at 12 VDC0.7 A or less
Current Consumption at 15 VDC0.6 A or less
Operating Temperature0 to 40°C
Operating Humidity20 to 85% RH, noncondensing
Input-to-Case Withstand Test250 VAC for 1 minute
Channel-to-Channel Withstand Test250 VAC for 1 minute
Protection RatingIP30, subject to input adapter configuration
Dimensions210 W × 35 H × 157.5 D mm, excluding protrusions
WeightApprox. 650 g
Applicable EU DirectivesRoHS and EMC
Specified Warm-Up TimeApproximately 1 hour
Standard USB CableN-38, 1 m
Standard Ground WireP-72, 5 m
Included SoftwareDCS-100A, DVD
Optional AC AdapterUNI360-1540-AC
Optional Stacking KitST-1B

Specifications are based on KYOWA's PCD-400B/430B Hardware Instruction Manual, IM-A-1438B (September 2023). The maximum 10 kHz sampling frequency is distinct from the DC to 200 Hz measurement frequency response. Specified accuracy applies after approximately one hour of warm-up and temperature stabilization.

The 250 VAC withstand values are insulation test conditions, not continuous permissible measurement input voltages. The PCD-400B measures strain gages and strain-gage transducers; voltage measurement mode is provided by the separate PCD-430B model. The external AC adapter, stacking accessories, and input adapters are optional items.

Model Description
ST-1BStacking kit for multiunit configurations
UI-10AInput adapter for strain-gage transducers; TEDS compatible
UI-11AInput adapter for strain gages; TEDS compatible
UI-15AStrain gage input adapter with operating levers
UI-16BOne-touch input adapter for strain gages
UI-55AOne-touch type input adapter
UNI360-1540-ACExternal AC power adapter
N-97Connection cable for compatible stacked configurations
DCS-100ADynamic data acquisition software
DAS-200AData analysis software
PCAS-100APCB strain and stress measuring software
PCAS-1000BPCB stress measuring set incorporating PCD-400B

Accessory and software availability may vary by region and package. Verify compatible input adapters, strain gage wiring, power supply, software licensing, and operating-system requirements before quotation.