4-Channel USB Sensor Interface for Strain Gage Measurement
The KYOWA PCD-400B is a four-channel USB sensor interface designed for PC-based measurement of strain gages and strain-gage transducers. It combines bridge excitation, signal conditioning, and 24-bit A/D conversion in a compact instrument for laboratory testing, product evaluation, and automated strain data acquisition.
Each PCD-400B provides four balanced differential input channels for compatible quarter-bridge, half-bridge, and full-bridge strain measurement configurations. With the optional stacking kit, up to four units can be combined to create a synchronous measurement system with a maximum of 16 channels.
The PCD-400B supports seven strain measurement ranges from ±200 to ±20,000 microstrain, with a specified measurement accuracy within ±0.5% of full scale. Its AC 2 Vrms bridge excitation and automatic resistance and capacitance balance adjustment simplify configuration of compatible strain measurement circuits.
KYOWA specifies a frequency response from DC to 200 Hz and a maximum sampling frequency of 10 kHz. These are separate characteristics: the 10 kHz sampling specification does not mean that the instrument has a 10 kHz analog measurement bandwidth.
USB 2.0 connectivity enables operation with compatible PC software, including KYOWA DCS-100A dynamic data acquisition software. Optional input adapters support strain gages and strain-gage transducers, while selected adapters provide TEDS sensor information reading.
The PCD-400B is also compatible with KYOWA PCAS-100A software for printed circuit board strain and stress evaluation. For a complete PCB stress measurement system, KYOWA offers the PCAS-1000B set combining the PCD-400B, dedicated software, and applicable accessories.
The PCD-400B is designed for strain measurement. Applications requiring a dedicated voltage measurement mode should consider the PCD-430B. A compatible DC power source and appropriate input adapters must be selected for the intended system.