SR-NIR Near-Infrared Spectroradiometer
The discontinued SR-NIR is a spot near-infrared spectroradiometer for evaluating weak spectral emission and transmission in the 600–1,030 nm region.
The SR-NIR uses a diffraction grating and electronically cooled linear CCD to analyze near-infrared spectral radiance from displays, optical films, and light sources. The instrument itself covers 600–1,030 nm. The frequently cited 380–1,030 nm range is possible only when the SR-NIR is used together with another visible-range SR series instrument.
Four motor-selectable observation angles and a 350 mm-to-infinity measurement distance help accommodate small optical targets. Published reference signal ranges by angle are based on standard illuminant A and must not be presented as a calibrated luminance measurement range; the manufacturer explicitly says the SR-NIR is not a brightness-measurement instrument.
The manufacturer lists the SR-NIR as discontinued without identifying a direct successor. For replacement projects, confirm the required spectral band, signal level, and integration method before selecting a current instrument.