World-Class Performance for Advanced Applications
• Sensor and furnace designs deliver low-noise, high-sensitivity, and excellent resolution to evaluate smallest of reactions. • A baseline repeatability of ±5 μW thanks to innovative furnace technologies for results you can trust, time after time. • Packaged with features like dual cooling and all software modules as standard including advanced DSC techniques to deliver excellent value for money.
• Intuitive software for ease of use and optional auto-sampler measuring up to 50 samples at once makes analysis easy and time efficient. • Built-in safety features, including an automated furnace cover with safety sensor, fans to ensure the outside doesn’t get too hot and an anti-overheating mechanism, safeguard operators when using the instrument. • Data analysis, record keeping, troubleshooting and reporting made easy to save time. • Our DSC instruments meet relevant standards including ISO, ASTM, DIN and JIS with ASTM E1269, ASTM E1641-16, ASTM E794-06, ASTM E793-06, ISO11357-4, ISO 11358-3, ISO 11357-3, ISO 11357-2 and ISO 22768 built into the software.
DSC200 | DSC600 | |
Heat Flow Measurement Method | Heat flux with centric heat flow | Heat flux with centric heat flow and 3 thermocouples |
Time Constant (Resolution) | 6.5 seconds or better | 3 seconds or better |
RMS Noise / Sensitivity | 0.1 µW / 0.2 µW | 0.05 µW / 0.1 µW |
Temperature Range | -150 to 725°C | |
Programmable Heating/Cooling Rate | 0.01 ~ 100°C/min | |
Temperature Precision | +/- 0.01°C | |
Temperature Accuracy | +/- 0.05°C | |
Calorimetric Precision | +/- 0.05% | |
Calorimetric Accuracy | +/- 0.4% | |
Baseline Stability | +/- 5 µW | |
Real View® Camera System | Optional | |